The fractionated dipole antenna: A new antenna for body imaging at 7 Tesla.
Raaijmakers, Alexander J E; Italiaander, Michel; Voogt, Ingmar J; Luijten, Peter R; Hoogduin, Johannes M; Klomp, Dennis W J; van den Berg, Cornelis A T.
Afiliação
  • Raaijmakers AJ; Department of Radiology, UMC Utrecht, Utrecht, The Netherlands.
  • Italiaander M; MR Coils B.V, Drunen, The Netherlands.
  • Voogt IJ; Department of Radiology, UMC Utrecht, Utrecht, The Netherlands.
  • Luijten PR; Department of Radiology, UMC Utrecht, Utrecht, The Netherlands.
  • Hoogduin JM; Department of Radiology, UMC Utrecht, Utrecht, The Netherlands.
  • Klomp DW; Department of Radiology, UMC Utrecht, Utrecht, The Netherlands.
  • van den Berg CA; Department of Radiotherapy, UMC Utrecht, Utrecht, The Netherlands.
Magn Reson Med ; 75(3): 1366-74, 2016 Mar.
Article em En | MEDLINE | ID: mdl-25939890
ABSTRACT

PURPOSE:

Dipole antennas in ultrahigh field MRI have demonstrated advantages over more conventional designs. In this study, the fractionated dipole antenna is presented a dipole where the legs are split into segments that are interconnected by capacitors or inductors.

METHODS:

A parameter study has been performed on dipole antenna length using numerical simulations. A subsequent simulation study investigates the optimal intersegment capacitor/inductor value. The resulting optimal design has been constructed and compared to a previous design, the single-side adapted dipole (SSAD) by simulations and measurements. An array of eight elements has been constructed for prostate imaging on four subjects (body mass index 20-27.5) using 8 × 2 kW amplifiers.

RESULTS:

For prostate imaging at 7T, lowest peak local specific-absorption rate (SAR) levels are achieved if the antenna is 30 cm or longer. A fractionated dipole antenna design with inductors between segments has been chosen to achieve even lower SAR levels and more homogeneous receive sensitivities.

CONCLUSION:

With the new design, good quality prostate images are acquired. SAR levels are reduced by 41% to 63% in comparison to the SSAD. Coupling levels are moderate (average nearest neighbor -14.6 dB) for each subject and prostate B1+ levels range from 12 to 18 µT.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética Limite: Adult / Humans / Male Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética Limite: Adult / Humans / Male Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Holanda