Evaluation of the PREVI® Isola automated seeder system compared to reference manual inoculation for antibiotic susceptibility testing by the disk diffusion method.
Le Page, S; van Belkum, A; Fulchiron, C; Huguet, R; Raoult, D; Rolain, J-M.
Afiliação
  • Le Page S; URMITE UM 63 CNRS 7278 IRD 198 INSERM U1905, IHU Méditerranée Infection, Faculté de Médecine et de Pharmacie, Aix-Marseille Université, 27 Boulevard Jean Moulin, 13385, Marseille CEDEX 05, France.
Eur J Clin Microbiol Infect Dis ; 34(9): 1859-69, 2015 Sep.
Article em En | MEDLINE | ID: mdl-26092031
The disk diffusion (DD) method remains the most popular manual technique for antibiotic susceptibility testing (AST) in clinical microbiology laboratories. This is because of its simplicity, reproducibility, and limited cost compared to (automated) microdilution systems, which are usually less sensitive at detecting certain important mechanisms of resistance. Here, we evaluate the PREVI® Isola automated seeder system using a new protocol for spreading bacterial suspensions (eight deposits of calibrated inocula of bacteria, followed by two rounds of rotation) in comparison with manual DD reference testing on a large series of clinical and reference strains. The average time required for seeding one agar plate for DD with this new protocol was 51 s per plate, i.e., 70 agar plates/h. Reproducibility and repeatability was assessed on three reference and three randomly chosen clinical strains, as usually requested by the European Committee on Antimicrobial Susceptibility Testing (EUCAST), and was excellent compared to the manual method. The standard deviations of zones of growth inhibition showed no statistical discrimination. The correlation between the two methods, assessed using 294 clinical isolates and a panel of six antibiotics (n = 3,528 zones of growth inhibition measured), was excellent, with a correlation coefficient of 0.977. The new PREVI® Isola protocol adapted for DD had a sensitivity of 99 % and a specificity of 100 % compared to the manual technique for interpreting DD as recommended by the EUCAST.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Bactérias / Testes de Sensibilidade a Antimicrobianos por Disco-Difusão / Automação Laboratorial Tipo de estudo: Diagnostic_studies / Evaluation_studies / Guideline Limite: Humans Idioma: En Revista: Eur J Clin Microbiol Infect Dis Assunto da revista: DOENCAS TRANSMISSIVEIS / MICROBIOLOGIA Ano de publicação: 2015 Tipo de documento: Article País de afiliação: França

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Bactérias / Testes de Sensibilidade a Antimicrobianos por Disco-Difusão / Automação Laboratorial Tipo de estudo: Diagnostic_studies / Evaluation_studies / Guideline Limite: Humans Idioma: En Revista: Eur J Clin Microbiol Infect Dis Assunto da revista: DOENCAS TRANSMISSIVEIS / MICROBIOLOGIA Ano de publicação: 2015 Tipo de documento: Article País de afiliação: França