Newborn manikin study shows that physicians often fail to detect correct lung compliance when using a self-inflating bag.
Boldingh, Anne Marthe; Solevåg, Anne Lee; Benth, Jurate Saltyte; Klingenberg, Claus; Nakstad, Britt.
Afiliação
  • Boldingh AM; Department of Paediatric and Adolescent Medicine, Akershus University Hospital, Akershus, Norway.
  • Solevåg AL; Institute of Clinical Medicine, Campus Akershus University Hospital, University of Oslo, Oslo, Norway.
  • Benth JS; Department of Paediatric and Adolescent Medicine, Akershus University Hospital, Akershus, Norway.
  • Klingenberg C; Institute of Clinical Medicine, Campus Akershus University Hospital, University of Oslo, Oslo, Norway.
  • Nakstad B; HØKH, Research Centre, Akershus University Hospital, Lørenskog, Norway.
Acta Paediatr ; 105(2): 172-7, 2016 Feb.
Article em En | MEDLINE | ID: mdl-26153507
ABSTRACT

AIM:

Recognising changes in lung compliance can help clinicians to adjust initial inflations during resuscitation at birth. We examined whether physicians sensed low and normal compliance with a self-inflating bag before and after an educational intervention that used a manikin connected to a newborn lung simulator.

METHODS:

We asked 43 physicians with neonatal duties to perform two low compliance ventilation attempts and two normal-compliance ventilation attempts in a randomised order at baseline and after the educational intervention, with 34 taking part in a retest three months later.

RESULTS:

The physicians correctly recognised low and normal compliance in 71% and 66% of the ventilations at baseline, 80% and 66% of the ventilations after the intervention and 74% and 81% at retest. Correct recognition of normal compliance improved from baseline to retest (p = 0.04). Ventilations in low- vs normal-compliance settings resulted in lower tidal volumes (4.4 vs 23.0 mL, p < 0.001), lower ventilation rates (42 vs 51, p < 0.001) and higher peak inflating pressure (35.2 vs 31.4 cmH2 O, p < 0.001).

CONCLUSION:

Around one in four physicians failed to recognise correct compliance levels when using a self-inflating bag and showed limited improvement after an educational intervention. Ventilations in a low-compliance setting resulted in suboptimal ventilation.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Complacência Pulmonar Tipo de estudo: Clinical_trials Limite: Adult / Female / Humans / Male / Newborn Idioma: En Revista: Acta Paediatr Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Noruega

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Complacência Pulmonar Tipo de estudo: Clinical_trials Limite: Adult / Female / Humans / Male / Newborn Idioma: En Revista: Acta Paediatr Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Noruega