Multidetector Computed Tomography Angiography Combined with Intravascular Ultrasound for the Evaluation of Coronary Veins for Percutaneous Mitral Annulus Repair Using Transcoronary Sinus Devices.
Chen, Shasha; Zhu, Qian; Ning, Zhongping; Wu, Yingbiao; Zhang, Zhen; Li, Xinming.
Afiliação
  • Chen S; Medical School, Tongji University, Shanghai, China.
  • Zhu Q; Cardiology Department, Zhoupu Hospital, Shanghai, China.
  • Ning Z; Cardiology Department, Zhoupu Hospital, Shanghai, China.
  • Wu Y; Cardiology Department, Zhoupu Hospital, Shanghai, China.
  • Zhang Z; Cardiology Department, Zhoupu Hospital, Shanghai, China.
  • Li X; Medical School, Tongji University, Shanghai, China.
Echocardiography ; 32(12): 1851-7, 2015 Dec.
Article em En | MEDLINE | ID: mdl-26178305
ABSTRACT

BACKGROUND:

Multidetector computed tomography angiography (MDCTA) can be used to evaluate the target location of transcoronary sinus devices. This study aimed to assess the accuracy of MDCTA in evaluating the target location of transcoronary sinus devices compared with intravascular ultrasound (IVUS). MATERIALS AND

METHODS:

Forty-two patients planned to undergo mitral valve repair (MVR) were prospectively enrolled at Zhoupu Hospital (China) including 15 with secondary mitral regurgitation (MR) grade ≥3. MDCTA was performed to measure the diameters of coronary sinus ostium (CSO) and proximal anterior interventricular vein (PAIV) and the distance between them.

RESULTS:

During MVR, these parameters were measured using IVUS before electrode insertion. There was a strong linear correlation between the diameter of CSO measured by MDCTA and IVUS (r = 0.967, P < 0.001), as well as for PAIV (r = 0.954, P < 0.001) and the distance between them (r = 0.986, P < 0.001). No significant differences were found between the results measured by MDCTA and IVUS. The patients with secondary MR grade ≥3 had significantly larger CSO and PAIV measured by IVUS (P = 0.003 and P = 0.017, respectively), as well as by MDCTA (P = 0.010 and P = 0.008, respectively).

CONCLUSION:

Dual-source MDCTA might allow the quantitative evaluation of the target location of transcoronary sinus devices with a good accuracy. It may be a good choice for guiding the selection of transcoronary sinus devices.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ultrassonografia de Intervenção / Cirurgia Assistida por Computador / Anuloplastia da Valva Mitral / Tomografia Computadorizada Multidetectores / Insuficiência da Valva Mitral Tipo de estudo: Diagnostic_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Echocardiography Assunto da revista: CARDIOLOGIA / DIAGNOSTICO POR IMAGEM Ano de publicação: 2015 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ultrassonografia de Intervenção / Cirurgia Assistida por Computador / Anuloplastia da Valva Mitral / Tomografia Computadorizada Multidetectores / Insuficiência da Valva Mitral Tipo de estudo: Diagnostic_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Echocardiography Assunto da revista: CARDIOLOGIA / DIAGNOSTICO POR IMAGEM Ano de publicação: 2015 Tipo de documento: Article País de afiliação: China