Large Metal Heads and Vitamin E Polyethylene Increase Frictional Torque in Total Hip Arthroplasty.
Meneghini, R Michael; Lovro, Luke R; Wallace, Joseph M; Ziemba-Davis, Mary.
Afiliação
  • Meneghini RM; Department of Orthopaedic Surgery, Indiana University School of Medicine, Fishers, Indiana; Orthopedics and Sports Medicine, Indiana University Health Physicians, Fishers, Indiana.
  • Lovro LR; Orthopedics and Sports Medicine, Indiana University Health Physicians, Fishers, Indiana.
  • Wallace JM; Department of Biomedical Engineering, Purdue School of Engineering and Technology at Indiana University-Purdue University, Indianapolis, Indiana.
  • Ziemba-Davis M; Orthopedics and Sports Medicine, Indiana University Health Physicians, Fishers, Indiana.
J Arthroplasty ; 31(3): 710-4, 2016 Mar.
Article em En | MEDLINE | ID: mdl-26482682
ABSTRACT

PURPOSE:

Trunnionosis has reemerged in modern total hip arthroplasty for reasons that remain unclear. Bearing frictional torque transmits forces to the modular head-neck interface, which may contribute to taper corrosion. The purpose of this study is to compare frictional torque of modern bearing couples in total hip arthroplasty.

METHODS:

Mechanical testing based on in vivo loading conditions was used to measure frictional torque. All bearing couples were lubricated and tested at 1 Hz for more than 2000 cycles. The bearing couples tested included conventional, highly crosslinked (XLPE) and vitamin E polyethylene, CoCr, and ceramic femoral heads and dual-mobility bearings. Statistical analysis was performed using Student t test for single-variable and analysis of variance for multivariant analysis. P ≤ .05 was considered statistically significant.

RESULTS:

Large CoCr metal heads (≥36 mm) substantially increased frictional torque against XLPE liners (P = .01), a finding not observed in ceramic heads. Vitamin E polyethylene substantially increased frictional torque compared with XLPE in CoCr and ceramic heads (P = .001), whereas a difference between conventional and XLPE was not observed (P = .69) with the numbers available. Dual-mobility bearing with ceramic inner head demonstrated the lowest mean frictional torque of all bearing couples.

CONCLUSION:

In this simulated in vivo model, large-diameter CoCr femoral heads and vitamin E polyethylene liners are associated with increased frictional torque compared with smaller metal heads and XLPE, respectively. The increased frictional torque of vitamin E polyethylene and larger-diameter femoral heads should be considered and further studied, along with reported benefits of these modern bearing couples.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Artroplastia de Quadril / Prótese de Quadril Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: J Arthroplasty Assunto da revista: ORTOPEDIA Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Artroplastia de Quadril / Prótese de Quadril Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: J Arthroplasty Assunto da revista: ORTOPEDIA Ano de publicação: 2016 Tipo de documento: Article