Your browser doesn't support javascript.
loading
High-resolution, reflection mode tomographic imaging. Part II: Application to echography.
Herment, A; Guglielmi, J P; Péronneau, P; Dumée, P.
Afiliação
  • Herment A; INSERM U. 256, Hôpital Broussais Paris, France.
Ultrason Imaging ; 11(1): 22-41, 1989 Jan.
Article em En | MEDLINE | ID: mdl-2660389
Principles of high-resolution, ultrasonic imaging using data acquisition by a compound scanning with a sector echograph are presented. The signal processing is based on both deconvolution and reflection mode tomography. Three of the methods that can be derived from these principles are selected due to their lower computation costs. Applications of these methods to synthetic data and test targets demonstrate that, with respect to 2D deconvolution, they offer: a gain in computation time of more than 8, an improvement in resolution of the order of 10 and an increase of S/N ratio of the order of 4. Finally, both the effects of limited acquisition angular window and of a variable propagation speed are illustrated.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia / Ultrassonografia Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Ultrason Imaging Ano de publicação: 1989 Tipo de documento: Article País de afiliação: França
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia / Ultrassonografia Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Ultrason Imaging Ano de publicação: 1989 Tipo de documento: Article País de afiliação: França