Establishing an accurate gas phase reference frequency to quantify 129 Xe chemical shifts in vivo.
Virgincar, Rohan S; Robertson, Scott H; Nouls, John; Degan, Simone; Schrank, Geoffry M; He, Mu; Driehuys, Bastiaan.
Afiliação
  • Virgincar RS; Center for In Vivo Microscopy, Duke University Medical Center, Durham, North Carolina, USA.
  • Robertson SH; Biomedical Engineering, Duke University, Durham, North Carolina, USA.
  • Nouls J; Center for In Vivo Microscopy, Duke University Medical Center, Durham, North Carolina, USA.
  • Degan S; Medical Physics Graduate Program, Duke University, Durham, North Carolina, USA.
  • Schrank GM; Center for In Vivo Microscopy, Duke University Medical Center, Durham, North Carolina, USA.
  • He M; Radiology, Duke University Medical Center, Durham, North Carolina, USA.
  • Driehuys B; Center for In Vivo Microscopy, Duke University Medical Center, Durham, North Carolina, USA.
Magn Reson Med ; 77(4): 1438-1445, 2017 04.
Article em En | MEDLINE | ID: mdl-27059646
ABSTRACT

PURPOSE:

129 Xe interacts with biological media to exhibit chemical shifts exceeding 200 ppm that report on physiology and pathology. Extracting this functional information requires shifts to be measured precisely. Historically, shifts have been reported relative to the gas-phase resonance originating from pulmonary airspaces. However, this frequency is not fixed-it is affected by bulk magnetic susceptibility, as well as Xe-N2 , Xe-Xe, and Xe-O2 interactions. In this study, we addressed this by introducing a robust method to determine the 0 ppm 129 Xe reference from in vivo data.

METHODS:

Respiratory-gated hyperpolarized 129 Xe spectra from the gas- and dissolved-phases were acquired in four mice at 2T from multiple axial slices within the thoracic cavity. Complex spectra were then fitted in the time domain to identify peaks.

RESULTS:

Gas-phase 129 Xe exhibited two distinct resonances corresponding to 129 Xe in conducting airways (varying from -0.6 ± 0.2 to 1.3 ± 0.3 ppm) and alveoli (relatively stable, at -2.2 ± 0.1 ppm). Dissolved-phase 129 Xe exhibited five reproducible resonances in the thorax at 198.4 ± 0.4, 195.5 ± 0.4, 193.9 ± 0.2, 191.3 ± 0.2, and 190.7 ± 0.3 ppm.

CONCLUSION:

The alveolar 129 Xe resonance exhibits a stable frequency across all mice. Therefore, it can provide a reliable in vivo reference frequency by which to characterize other spectroscopic shifts. Magn Reson Med 771438-1445, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Alvéolos Pulmonares / Isótopos de Xenônio / Espectroscopia de Ressonância Magnética Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Animals Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Alvéolos Pulmonares / Isótopos de Xenônio / Espectroscopia de Ressonância Magnética Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Animals Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Estados Unidos