Microbore liquid chromatography and refractive index gradient detection of low-nanogram and low-ppm quantities of carbohydrates.
Hancock, D O; Synovec, R E.
Afiliação
  • Hancock DO; Department of Chemistry, University of Washington, Seattle 98195.
J Chromatogr ; 464(1): 83-91, 1989 Feb 17.
Article em En | MEDLINE | ID: mdl-2715252
A refractive index gradient detector is presented as a universal detector in the microbore high-performance liquid chromatography analysis of carbohydrates. Simultaneously, low-ng and low-ppm injected quantities of carbohydrates were detected at the 3 x root-mean-square baseline noise level. A typical microbore high-performance liquid chromatography chromatogram separating fructose from sucrose followed by refractive index gradient (RIG) detection is reported. Use of a position sensitive detector (PSD) in the RIG detector design is reported and experimental considerations discussed. Optimization of the PSD-based RIG detector is addressed. Potential for the device in industrial and clinical applications is considered.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Carboidratos Tipo de estudo: Diagnostic_studies Idioma: En Revista: J Chromatogr Ano de publicação: 1989 Tipo de documento: Article
Buscar no Google