Wen, B C; Pennington, E C; Hussey, D H; Jani, S K.
Afiliação
  • Wen BC; Department of Radiology, University of Iowa College of Medicine, Iowa City 52242.
Int J Radiat Oncol Biol Phys ; 16(6): 1637-41, 1989 Jun.
Article em En | MEDLINE | ID: mdl-2722601
ABSTRACT
Excessive irradiation due to unexpected leakage was found on a patient receiving electron beam therapy. The cause of this leakage was analyzed and the amount of leakage was measured for different electron beam energies. The highest leakage occurred with a 6 x 6 cm cone using a 12 MeV electron beam. The leakage dose measured along the side of the cone could be as great as 40%. Until the cones are modified or redesigned, it is advised that all patient setups be carefully reviewed to assure that no significant patient areas are in the side scatter region.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Radioterapia / Elétrons / Alopecia Tipo de estudo: Risk_factors_studies Limite: Humans / Male / Middle aged Idioma: En Revista: Int J Radiat Oncol Biol Phys Ano de publicação: 1989 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Radioterapia / Elétrons / Alopecia Tipo de estudo: Risk_factors_studies Limite: Humans / Male / Middle aged Idioma: En Revista: Int J Radiat Oncol Biol Phys Ano de publicação: 1989 Tipo de documento: Article