Alopecia associated with unexpected leakage from electron cone.
Int J Radiat Oncol Biol Phys
; 16(6): 1637-41, 1989 Jun.
Article
em En
| MEDLINE
| ID: mdl-2722601
ABSTRACT
Excessive irradiation due to unexpected leakage was found on a patient receiving electron beam therapy. The cause of this leakage was analyzed and the amount of leakage was measured for different electron beam energies. The highest leakage occurred with a 6 x 6 cm cone using a 12 MeV electron beam. The leakage dose measured along the side of the cone could be as great as 40%. Until the cones are modified or redesigned, it is advised that all patient setups be carefully reviewed to assure that no significant patient areas are in the side scatter region.