Your browser doesn't support javascript.
loading
[Possible reasons for cerebrospinal fluid gusher in cochlear implantation with inner ear abnormality].
Sun, S P; Lu, W; Men, X M; Zuo, B; Lei, Y B.
Afiliação
  • Sun SP; Division of Otology, Otorhinolaryngology Hospital, the First Affiliated Hospital of Zhengzhou University, Henan 450052, China.
  • Lu W; Division of Otology, Otorhinolaryngology Hospital, the First Affiliated Hospital of Zhengzhou University, Henan 450052, China.
  • Men XM; Division of Otology, Otorhinolaryngology Hospital, the First Affiliated Hospital of Zhengzhou University, Henan 450052, China.
  • Zuo B; Division of Otology, Otorhinolaryngology Hospital, the First Affiliated Hospital of Zhengzhou University, Henan 450052, China.
  • Lei YB; Division of Otology, Otorhinolaryngology Hospital, the First Affiliated Hospital of Zhengzhou University, Henan 450052, China.
Article em Zh | MEDLINE | ID: mdl-28441806
ABSTRACT

Objective:

To discuss the possible reasons for cerebrospinal fluid (CSF) gusher in cochlear implantation (CI) with inner ear abnormality.

Method:

A retrospective analysis was performed on 340 cases who underwent CI from January 2013 to December 2016 in Division of Otology, Otorhinolaryngology Hospital, the First Affiliated Hospital of Zhengzhou University. Among them, 96 cases had inner ear abnormalities. Imaging examinations were performed on these patients, and classification of inner ear malformation was done according to the results.

Results:

Among the cases with inner ear abnormality, 9.4% (9/96) suffered from CSF gusher during CI. The inner ear abnormalities were found to be as follows 3 cases had incomplete partition type Ⅰ; 1 case had incomplete partition type Ⅰ with semicircular canal dysplasia; 1 case had common cavity deformity; 1 case had enlarged vestibular aqueducts and common cavity deformity; 2 cases had Mondini deformity. All of these cases had bony defect in the fundus of the internal acoustic meatus observed on CT scans. Another case was type 1 cochlear aqueduct with round window aplasia.

Conclusions:

Defects in the modiolus or fundus of the internal acoustic meatus is the main reason for CSF gusher during CI. A patent cochlear aqueduct is another possible reason.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Osso Temporal / Aqueduto Vestibular / Canais Semicirculares / Otorreia de Líquido Cefalorraquidiano / Implante Coclear / Perda Auditiva Neurossensorial / Orelha Interna Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Child / Female / Humans / Male Idioma: Zh Revista: Zhonghua Er Bi Yan Hou Tou Jing Wai Ke Za Zhi Ano de publicação: 2017 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Osso Temporal / Aqueduto Vestibular / Canais Semicirculares / Otorreia de Líquido Cefalorraquidiano / Implante Coclear / Perda Auditiva Neurossensorial / Orelha Interna Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Child / Female / Humans / Male Idioma: Zh Revista: Zhonghua Er Bi Yan Hou Tou Jing Wai Ke Za Zhi Ano de publicação: 2017 Tipo de documento: Article País de afiliação: China