Microstructural alterations owing to handling of bovine pericardium to manufacture bioprosthetic heart valves: A potential risk for cusp dehiscence.
Mao, J; Wang, Y; Philippe, E; Cianciulli, T; Vesely, I; How, D; Bourget, J-M; Germain, L; Zhang, Z; Guidoin, R.
Afiliação
  • Mao J; Department of Surgery, Faculty of Medicine, Laval University and Axe médecine régénérative, centre de recherches du CHU, Ferdinand-Vandry Building, Room 4873, 2325, rue de l'Université, Québec G1V 0A6, QC, Canada.
  • Wang Y; Department of Surgery, Faculty of Medicine, Laval University and Axe médecine régénérative, centre de recherches du CHU, Ferdinand-Vandry Building, Room 4873, 2325, rue de l'Université, Québec G1V 0A6, QC, Canada.
  • Philippe E; Department of Surgery, Faculty of Medicine, Laval University and Axe médecine régénérative, centre de recherches du CHU, Ferdinand-Vandry Building, Room 4873, 2325, rue de l'Université, Québec G1V 0A6, QC, Canada.
  • Cianciulli T; Department of Cardiology, Hospital of the Government of the City of Buenos Aires "Prof. Cosme Argerich", Pi y Margall 780, C1155AHB Ciudad Autónoma de Buenos Aires, Argentina.
  • Vesely I; Cardiology, School of Medicine, University of Colorado Denver, Denver (CO) 80045, USA.
  • How D; Peninsula School of Medicine and Dentistry, Plymouth University, PL4 8AA, Plymouth, UK.
  • Bourget JM; Department of Surgery, Faculty of Medicine, Laval University and Axe médecine régénérative, centre de recherches du CHU, Ferdinand-Vandry Building, Room 4873, 2325, rue de l'Université, Québec G1V 0A6, QC, Canada.
  • Germain L; Department of Surgery, Faculty of Medicine, Laval University and Axe médecine régénérative, centre de recherches du CHU, Ferdinand-Vandry Building, Room 4873, 2325, rue de l'Université, Québec G1V 0A6, QC, Canada.
  • Zhang Z; Department of Surgery, Faculty of Medicine, Laval University and Axe médecine régénérative, centre de recherches du CHU, Ferdinand-Vandry Building, Room 4873, 2325, rue de l'Université, Québec G1V 0A6, QC, Canada.
  • Guidoin R; Department of Surgery, Faculty of Medicine, Laval University and Axe médecine régénérative, centre de recherches du CHU, Ferdinand-Vandry Building, Room 4873, 2325, rue de l'Université, Québec G1V 0A6, QC, Canada. Electronic address: robert.guidoin@fmed.ulaval.ca.
Morphologie ; 101(333): 77-87, 2017 Jun.
Article em En | MEDLINE | ID: mdl-28442174
ABSTRACT

INTRODUCTION:

Cross-linking and anti-calcification of prosthetic heart valves have been continuously improved to prevent degeneration and calcification. However, non-calcific structural deteriorations such as cuspal dehiscences along the stent still require further analysis. MATERIAL AND

METHOD:

Based upon the previous analysis of an explanted valve after 7 years, a fresh commercial aortic valve was embedded in poly(methyl methacrylate) (PMMA) and cut into slices to ensure the detailed observation of the assembly and material structures. A pericardial patch embossed to provide the adequate shape of the cusps was investigated after paraffin embedding and appropriate staining. The microstructural damages that occurred during manufacturing process were identified and evaluated by light microscopy, polarized microscopy, scanning electron microscopy (SEM) and transmission electron microscopy (TEM).

RESULTS:

The wavy collagen bundles, the key structure of the pericardium patch, were damaged to a great extent at suture sites along the stent and in the compressed areas around the stent post. The fixation of the embossed pericardium patch along the plots of the stent aggravated the microstructural modifications. The damages mainly appeared as the elimination of collagen bundle waviness and delamination between the bundles.

CONCLUSION:

Considering the modes of failure of the explant, the damages to the collagen bundles may identify the vulnerable sites that play an important role in the cusp dehiscence of heart valve implants. Such information is important to the manufacturers. Recommendations to prevent in vivo cusp dehiscence can therefore be formulated.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Pericárdio / Manejo de Espécimes / Bioprótese / Próteses Valvulares Cardíacas Tipo de estudo: Etiology_studies / Risk_factors_studies Limite: Animals Idioma: En Revista: Morphologie Assunto da revista: ANATOMIA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Canadá

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Pericárdio / Manejo de Espécimes / Bioprótese / Próteses Valvulares Cardíacas Tipo de estudo: Etiology_studies / Risk_factors_studies Limite: Animals Idioma: En Revista: Morphologie Assunto da revista: ANATOMIA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Canadá