Systematic influences on the areas of peaks in gamma-ray spectra that have a large statistical uncertainty.
Bruggeman, M; Collins, S M; Done, L; Durasevic, M; Duch, M A; Gudelis, A; Hyza, M; Jevremovic, A; Kandic, A; Korun, M; Ilie, S; Lee, J M; Lee, K B; Luca, A; Margineanu, R M; Pantelica, A; Serrano, I; Seslak, B; Tugulan, L C; Verheyen, L; Vodenik, B; Vukanac, I; Zeng, Z; Zorko, B.
Afiliação
  • Bruggeman M; Studiecentrum voor Kernenergie, Kernenergie, Boeretang 200, B-2400 Mol, Belgium.
  • Collins SM; National Physical Laboratory, Teddington, Middlesex TW11 0LW, UK.
  • Done L; Horia Hulubei National Institute for Research and Development in Physics and Nuclear Engineering (IFIN_HH), 30 Reactorului Street, POB MG-6, RO-077125 Bucharest-Magurele, Romania.
  • Durasevic M; Institute of Nuclear Sciences "Vinca", Laboratory for Nuclear and Plasma Physics, University of Belgrade, Belgrade, Serbia.
  • Duch MA; Universitat Politècnica de Catalunya (UPC), Institut de Tècniques Energetiques, Diagonal 647, 08028 Barcelona, Spain.
  • Gudelis A; Center for Physical Sciences and Technology, Savanoriu Ave. 231, Vilnus, Lithuania.
  • Hyza M; National Radiation Protection Institute, Bartoskova 1450/28, 140 00 Praha 4, Czech Republic.
  • Jevremovic A; Institute of Nuclear Sciences "Vinca", Laboratory for Nuclear and Plasma Physics, University of Belgrade, Belgrade, Serbia.
  • Kandic A; Institute of Nuclear Sciences "Vinca", Laboratory for Nuclear and Plasma Physics, University of Belgrade, Belgrade, Serbia.
  • Korun M; "Jozef Stefan" Institute, Jamova cesta 39, 1000 Ljubljana, Slovenia. Electronic address: matjaz.korun@ijs.si.
  • Ilie S; Horia Hulubei National Institute for Research and Development in Physics and Nuclear Engineering (IFIN_HH), 30 Reactorului Street, POB MG-6, RO-077125 Bucharest-Magurele, Romania.
  • Lee JM; Korea Research Institute of Standards and Science, Daejeon 305-340, Republic of Korea.
  • Lee KB; Korea Research Institute of Standards and Science, Daejeon 305-340, Republic of Korea.
  • Luca A; Horia Hulubei National Institute for Research and Development in Physics and Nuclear Engineering (IFIN_HH), 30 Reactorului Street, POB MG-6, RO-077125 Bucharest-Magurele, Romania.
  • Margineanu RM; Horia Hulubei National Institute for Research and Development in Physics and Nuclear Engineering (IFIN_HH), 30 Reactorului Street, POB MG-6, RO-077125 Bucharest-Magurele, Romania.
  • Pantelica A; Horia Hulubei National Institute for Research and Development in Physics and Nuclear Engineering (IFIN_HH), 30 Reactorului Street, POB MG-6, RO-077125 Bucharest-Magurele, Romania.
  • Serrano I; Universitat Politècnica de Catalunya (UPC), Institut de Tècniques Energetiques, Diagonal 647, 08028 Barcelona, Spain.
  • Seslak B; Institute of Nuclear Sciences "Vinca", Laboratory for Nuclear and Plasma Physics, University of Belgrade, Belgrade, Serbia.
  • Tugulan LC; Horia Hulubei National Institute for Research and Development in Physics and Nuclear Engineering (IFIN_HH), 30 Reactorului Street, POB MG-6, RO-077125 Bucharest-Magurele, Romania.
  • Verheyen L; Studiecentrum voor Kernenergie, Kernenergie, Boeretang 200, B-2400 Mol, Belgium.
  • Vodenik B; "Jozef Stefan" Institute, Jamova cesta 39, 1000 Ljubljana, Slovenia.
  • Vukanac I; Institute of Nuclear Sciences "Vinca", Laboratory for Nuclear and Plasma Physics, University of Belgrade, Belgrade, Serbia.
  • Zeng Z; Tsinghua University, Department of Engineering Physics, Beijing, PR China.
  • Zorko B; "Jozef Stefan" Institute, Jamova cesta 39, 1000 Ljubljana, Slovenia.
Appl Radiat Isot ; 134: 51-55, 2018 Apr.
Article em En | MEDLINE | ID: mdl-28673731
ABSTRACT
A method is presented for calculating the expected number of counts in peaks that have a large relative peak-area uncertainty and appear in measured gamma-ray spectra. The method was applied to calculations of the correction factors for peaks occurring in the spectra of radon daughters. It was shown that the factors used for correcting the calculated peak areas to their expected values decrease with an increasing relative peak-area uncertainty. The accuracy of taking the systematic influence inducing the correction factors into account is given by the dispersion of the correction factors corresponding to specific peaks. It was shown that the highest accuracy is obtained in the peak analyses with the GammaVision and Gamma-W software.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Appl Radiat Isot Assunto da revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Bélgica

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Appl Radiat Isot Assunto da revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Bélgica