EVALUATION OF A NEW ANAESTHESIA MACHINE (IN LOW FLOW CONDITION).
Bhargava, A K; Mahajan, H K; Gandhi, B K; Sharma, S.
Afiliação
  • Bhargava AK; Classified Specialist (Anaesthesiology), Department of Anaesthesiology, Command Hospital (NC), C/o 56 APO.
  • Mahajan HK; Graded Specialist (Anaesthesiology), Department of Anaesthesiology, Command Hospital (NC), C/o 56 APO.
  • Gandhi BK; MO Anaesthesiologist, Department of Anaesthesiology, Command Hospital (NC), C/o 56 APO.
  • Sharma S; Classified Specialist (Anaesthesiology), Department of Anaesthesiology, Command Hospital (NC), C/o 56 APO.
Med J Armed Forces India ; 53(4): 259-262, 1997 Oct.
Article em En | MEDLINE | ID: mdl-28769507
ABSTRACT
The performance of a new anaesthesia machine was evaluated under low flow anaesthetic technique for 7 months. An assessment under low flow conditions over a definitive period can highlight the drawbacks or advantages of a machine under stringent conditions. The overall performance was satisfactory in the sense that it proved to be an efficient, reliable and safe machine. Minimum flows of oxygen and nitrous oxide that could be achieved at steady state were 0.4 L/min and 0.5 L/min respectively. At these flows the machine monitors were accurate and one could safely administer anaesthesia with a significant amount of saving of nitrous oxide. The requirement of nitrous oxide fell from 2 cylinders in a day to 2 cylinders in a week.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Med J Armed Forces India Ano de publicação: 1997 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Med J Armed Forces India Ano de publicação: 1997 Tipo de documento: Article