Using Patient Feedback to Optimize the Design of a Certolizumab Pegol Electromechanical Self-Injection Device: Insights from Human Factors Studies.
Domanska, Barbara; Stumpp, Oliver; Poon, Steven; Oray, Serkan; Mountian, Irina; Pichon, Clovis.
Afiliação
  • Domanska B; UCB Pharma, Slough, UK. barbara.domanska@ucb.com.
  • Stumpp O; UCB Pharma, Bulle, Switzerland.
  • Poon S; UCB Pharma, Slough, UK.
  • Oray S; UCB Pharma, Braine-l'Alleud, Belgium.
  • Mountian I; UCB Pharma, Brussels, Belgium.
  • Pichon C; UCB Pharma, Bulle, Switzerland.
Adv Ther ; 35(1): 100-115, 2018 01.
Article em En | MEDLINE | ID: mdl-29222625
ABSTRACT

INTRODUCTION:

We incorporated patient feedback from human factors studies (HFS) in the patient-centric design and validation of ava®, an electromechanical device (e-Device) for self-injecting the anti-tumor necrosis factor certolizumab pegol (CZP).

METHODS:

Healthcare professionals, caregivers, healthy volunteers, and patients with rheumatoid arthritis, psoriatic arthritis, ankylosing spondylitis, or Crohn's disease participated in 11 formative HFS to optimize the e-Device design through intended user feedback; nine studies involved simulated injections. Formative participant questionnaire feedback was collected following e-Device prototype handling. Validation HFS (one EU study and one US study) assessed the safe and effective setup and use of the e-Device using 22 predefined critical tasks. Task outcomes were categorized as "failures" if participants did not succeed within three attempts.

RESULTS:

Two hundred eighty-three participants entered formative (163) and validation (120) HFS; 260 participants performed one or more simulated e-Device self-injections. Design changes following formative HFS included alterations to buttons and the graphical user interface screen. All validation HFS participants completed critical tasks necessary for CZP dose delivery, with minimal critical task failures (12 of 572 critical tasks, 2.1%, in the EU study, and 2 of 5310 critical tasks, less than 0.1%, in the US study).

CONCLUSION:

CZP e-Device development was guided by intended user feedback through HFS, ensuring the final design addressed patients' needs. In both validation studies, participants successfully performed all critical tasks, demonstrating safe and effective e-Device self-injections.

FUNDING:

UCB Pharma. Plain language summary available on the journal website.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artrite / Doença de Crohn / Antirreumáticos / Desenho de Equipamento / Certolizumab Pegol Tipo de estudo: Qualitative_research Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Adv Ther Assunto da revista: TERAPEUTICA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artrite / Doença de Crohn / Antirreumáticos / Desenho de Equipamento / Certolizumab Pegol Tipo de estudo: Qualitative_research Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Adv Ther Assunto da revista: TERAPEUTICA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Reino Unido