Your browser doesn't support javascript.
loading
Small field dosimetry for the small animal radiotherapy research platform (SARRP).
Ghita, Mihaela; McMahon, Stephen J; Thompson, Hannah F; McGarry, Conor K; King, Raymond; Osman, Sarah O S; Kane, Jonathan L; Tulk, Amanda; Schettino, Giuseppe; Butterworth, Karl T; Hounsell, Alan R; Prise, Kevin M.
Afiliação
  • Ghita M; Centre for Cancer Research and Cell Biology, Queen's University Belfast, 97 Lisburn Road, Belfast, BT9 7AE, UK. m.ghita@qub.ac.uk.
  • McMahon SJ; Centre for Cancer Research and Cell Biology, Queen's University Belfast, 97 Lisburn Road, Belfast, BT9 7AE, UK.
  • Thompson HF; Centre for Cancer Research and Cell Biology, Queen's University Belfast, 97 Lisburn Road, Belfast, BT9 7AE, UK.
  • McGarry CK; Radiotherapy Physics, Northern Ireland Cancer Centre, Belfast City Hospital, Belfast, BT9 7AB, UK.
  • King R; Centre for Cancer Research and Cell Biology, Queen's University Belfast, 97 Lisburn Road, Belfast, BT9 7AE, UK.
  • Osman SOS; Radiotherapy Physics, Northern Ireland Cancer Centre, Belfast City Hospital, Belfast, BT9 7AB, UK.
  • Kane JL; Centre for Cancer Research and Cell Biology, Queen's University Belfast, 97 Lisburn Road, Belfast, BT9 7AE, UK.
  • Tulk A; Radiotherapy Physics, Northern Ireland Cancer Centre, Belfast City Hospital, Belfast, BT9 7AB, UK.
  • Schettino G; Xstrahl Inc, 480 Brogdon Road, Suite 300, Suwanee, GA, 30024, USA.
  • Butterworth KT; Xstrahl Ltd, The Coliseum, Watchmoor Park, Riverside Way, Camberley, Surrey, GU15 3YL, UK.
  • Hounsell AR; National Physical Laboratory, Hampton Road, Teddington, Middlesex, TW11 0LW, UK.
  • Prise KM; Centre for Cancer Research and Cell Biology, Queen's University Belfast, 97 Lisburn Road, Belfast, BT9 7AE, UK.
Radiat Oncol ; 12(1): 204, 2017 Dec 28.
Article em En | MEDLINE | ID: mdl-29282134
ABSTRACT

BACKGROUND:

Preclinical radiation biology has become increasingly sophisticated due to the implementation of advanced small animal image guided radiation platforms into laboratory investigation. These small animal radiotherapy devices enable state-of-the-art image guided therapy (IGRT) research to be performed by combining high-resolution cone beam computed tomography (CBCT) imaging with an isocentric irradiation system. Such platforms are capable of replicating modern clinical systems similar to those that integrate a linear accelerator with on-board CBCT image guidance.

METHODS:

In this study, we present a dosimetric evaluation of the small animal radiotherapy research platform (SARRP, Xstrahl Inc.) focusing on small field dosimetry. Physical dosimetry was assessed using ion chamber for calibration and radiochromic film, investigating the impact of beam focus size on the dose rate output as well as beam characteristics (beam shape and penumbra). Two film analysis tools) have been used to assess the dose output using the 0.5 mm diameter aperture.

RESULTS:

Good agreement (between 1.7-3%) was found between the measured physical doses and the data provided by Xstrahl for all apertures used. Furthermore, all small field dosimetry data are in good agreement for both film reading methods and with our Monte Carlo simulations for both focal spot sizes. Furthermore, the small focal spot has been shown to produce a more homogenous beam with more stable penumbra over time.

CONCLUSIONS:

FilmQA Pro is a suitable tool for small field dosimetry, with a sufficiently small sampling area (0.1 mm) to ensure an accurate measurement. The electron beam focus should be chosen with care as this can potentially impact on beam stability and reproducibility.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiobiologia / Planejamento da Radioterapia Assistida por Computador / Imagens de Fantasmas / Tomografia Computadorizada de Feixe Cônico Tipo de estudo: Guideline / Health_economic_evaluation Limite: Animals Idioma: En Revista: Radiat Oncol Assunto da revista: NEOPLASIAS / RADIOTERAPIA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiobiologia / Planejamento da Radioterapia Assistida por Computador / Imagens de Fantasmas / Tomografia Computadorizada de Feixe Cônico Tipo de estudo: Guideline / Health_economic_evaluation Limite: Animals Idioma: En Revista: Radiat Oncol Assunto da revista: NEOPLASIAS / RADIOTERAPIA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Reino Unido