High bacterial contamination rate of electrocautery tips during total hip and knee arthroplasty.
Abdelaziz, Hussein; Zahar, Akos; Lausmann, Christian; Gehrke, Thorsten; Fickenscher, Helmut; Suero, Eduardo M; Gebauer, Matthias; Citak, Mustafa.
Afiliação
  • Abdelaziz H; Department of Orthopaedic Surgery, Helios ENDO-Klinik, Holstenstr. 2, 22767, Hamburg, Germany.
  • Zahar A; Department of Orthopaedic Surgery, Helios ENDO-Klinik, Holstenstr. 2, 22767, Hamburg, Germany.
  • Lausmann C; Department of Orthopaedic Surgery, Helios ENDO-Klinik, Holstenstr. 2, 22767, Hamburg, Germany.
  • Gehrke T; Department of Orthopaedic Surgery, Helios ENDO-Klinik, Holstenstr. 2, 22767, Hamburg, Germany.
  • Fickenscher H; Institute for Infection Medicine, Kiel University, Kiel, Germany.
  • Suero EM; Trauma Department, Hannover Medical School, Hannover, Germany.
  • Gebauer M; Department of Orthopaedic Surgery, Roland-Klinik Bremen, Bremen, Germany.
  • Citak M; Department of Orthopaedic Surgery, Helios ENDO-Klinik, Holstenstr. 2, 22767, Hamburg, Germany. mcitak@gmx.de.
Int Orthop ; 42(4): 755-760, 2018 04.
Article em En | MEDLINE | ID: mdl-29435622
ABSTRACT

PURPOSE:

The aim of the study was to quantify the bacterial contamination rate of electrocautery tips during primary total joint replacement (TJR), as well as during aseptic and septic revision TJR.

METHODS:

A total of 150 electrocautery tips were collected between April and July 2017. TJR surgeries were divided into three groups (1) primary, (2) aseptic and (3) septic revisions. In each group, a total of 50 electrocautery tips were collected. A monopolar electrocautery with a reusable stainless-steel blade tip was used in all cases. The rate of bacterial contamination was determined for all groups. Correlation of exposure time and type of surgery was analyzed.

RESULTS:

The overall bacterial contamination rate was 14.7% (95% CI 9.4 to 21.4%). The highest contamination rate occurred in the septic revision group (30.0%; 95% CI 17.9 to 44.6%), followed by the primary cases group (10.0%; 95% CI 3.3 to 21.8%) and the aseptic revision group (4.0%; 95% CI 0.5 to 13.7%). Exposure time did not affect the bacterial contamination rate. In 12 out of 15 (80%) contaminations identified in the septic group, we found the same causative microorganism of the prosthetic joint infection on the electrocautery tip.

CONCLUSIONS:

The bacterial contamination of the electrocautery tips is relatively high, especially during septic hip revision arthroplasty. Electrocautery tips should be changed after debridement of infected tissue.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Reoperação / Contaminação de Equipamentos / Artroplastia de Quadril / Artroplastia do Joelho / Eletrocoagulação Tipo de estudo: Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Int Orthop Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Reoperação / Contaminação de Equipamentos / Artroplastia de Quadril / Artroplastia do Joelho / Eletrocoagulação Tipo de estudo: Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Int Orthop Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha