Your browser doesn't support javascript.
loading
Ultrasound detection at fiber end-facets with surface plasmon resonance cavities.
Opt Lett ; 43(4): 775-778, 2018 Feb 15.
Article em En | MEDLINE | ID: mdl-29443991
Ultrasound detection is performed by measuring laser reflection off a surface plasmon resonance cavity which is integrated at a single-mode fiber end facet. It shows a total noise (or laser RIN) equivalent pressure of 9.7 KPa (or 5.2 KPa) over 0-20 MHz, a 6-dB angular detection range as large as 70° near 10 MHz, a detection bandwidth larger than 125 MHz, and a stable performance for over 20 min without feedback in ambient conditions. Its small form factor, fiber-optic integration, almost omnidirectional responsivity and large bandwidth are favorable for in vivo applications and high resolution imaging.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Revista: Opt Lett Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Revista: Opt Lett Ano de publicação: 2018 Tipo de documento: Article