Your browser doesn't support javascript.
loading
Quantification of Intracranial Aneurysm Volume Pulsation with 7T MRI.
Kleinloog, R; Zwanenburg, J J M; Schermers, B; Krikken, E; Ruigrok, Y M; Luijten, P R; Visser, F; Regli, L; Rinkel, G J E; Verweij, B H.
Afiliação
  • Kleinloog R; From the Department of Neurology and Neurosurgery (R.K., B.S., E.K., Y.M.R., L.R., G.J.E.R., B.H.V.), Brain Center Rudolf Magnus rachelkleinloog@gmail.com.
  • Zwanenburg JJM; Department of Radiology (J.J.M.Z., P.R.L., F.V.), Utrecht University, University Medical Center Utrecht, Utrecht, the Netherlands.
  • Schermers B; From the Department of Neurology and Neurosurgery (R.K., B.S., E.K., Y.M.R., L.R., G.J.E.R., B.H.V.), Brain Center Rudolf Magnus.
  • Krikken E; Department of Technical Medicine (B.S., E.K.), Faculty of Science and Technology, University of Twente, Enschede, the Netherlands.
  • Ruigrok YM; From the Department of Neurology and Neurosurgery (R.K., B.S., E.K., Y.M.R., L.R., G.J.E.R., B.H.V.), Brain Center Rudolf Magnus.
  • Luijten PR; Department of Technical Medicine (B.S., E.K.), Faculty of Science and Technology, University of Twente, Enschede, the Netherlands.
  • Visser F; From the Department of Neurology and Neurosurgery (R.K., B.S., E.K., Y.M.R., L.R., G.J.E.R., B.H.V.), Brain Center Rudolf Magnus.
  • Regli L; Department of Radiology (J.J.M.Z., P.R.L., F.V.), Utrecht University, University Medical Center Utrecht, Utrecht, the Netherlands.
  • Rinkel GJE; Department of Radiology (J.J.M.Z., P.R.L., F.V.), Utrecht University, University Medical Center Utrecht, Utrecht, the Netherlands.
  • Verweij BH; Philips Healthcare (F.V.), Best, the Netherlands.
AJNR Am J Neuroradiol ; 39(4): 713-719, 2018 Apr.
Article em En | MEDLINE | ID: mdl-29472302
BACKGROUND AND PURPOSE: Aneurysm volume pulsation is a potential predictor of intracranial aneurysm rupture. We evaluated whether 7T MR imaging can quantify aneurysm volume pulsation. MATERIALS AND METHODS: In Stage I of the study, 10 unruptured aneurysms in 9 patients were studied using a high-resolution (0.6-mm, isotropic) 3D gradient-echo sequence with cardiac gating. Semiautomatic segmentation was used to measure aneurysm volume (in cubic millimeters) per cardiac phase. Aneurysm pulsation was defined as the relative increase in volume between the phase with the smallest volume and the phase with the largest volume. The accuracy and precision of the measured volume pulsations were addressed by digital phantom simulations and a repeat image analysis. In Stage II, the imaging protocol was optimized and 9 patients with 9 aneurysms were studied with and without administration of a contrast agent. RESULTS: The mean aneurysm pulsation in Stage I was 8% ± 7% (range, 2%-27%), with a mean volume change of 15 ± 14 mm3 (range, 3-51 mm3). The mean difference in volume change for the repeat image analysis was 2 ± 6 mm3. The artifactual volume pulsations measured with the digital phantom simulations were of the same magnitude as the volume pulsations observed in the patient data, even after protocol optimization in Stage II. CONCLUSIONS: Volume pulsation quantification with the current imaging protocol on 7T MR imaging is not accurate due to multiple imaging artifacts. Future studies should always include aneurysm-specific accuracy analysis.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Aneurisma Intracraniano / Neuroimagem Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: AJNR Am J Neuroradiol Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Aneurisma Intracraniano / Neuroimagem Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: AJNR Am J Neuroradiol Ano de publicação: 2018 Tipo de documento: Article