Shorter treatment times reduce the impact of intra-fractional motion : A real-time 4DUS study comparing VMAT vs. step-and-shoot IMRT for prostate cancer.
Ballhausen, Hendrik; Li, Minglun; Ganswindt, Ute; Belka, Claus.
Afiliação
  • Ballhausen H; Department of Radiation Oncology, University Hospital, LMU Munich, Marchioninistraße 15, 81377, Munich, Germany. Hendrik.Ballhausen@med.uni-muenchen.de.
  • Li M; Department of Radiation Oncology, University Hospital, LMU Munich, Marchioninistraße 15, 81377, Munich, Germany.
  • Ganswindt U; Department of Radiation Oncology, University Hospital, LMU Munich, Marchioninistraße 15, 81377, Munich, Germany.
  • Belka C; Department of Radiation Oncology, University Hospital, LMU Munich, Marchioninistraße 15, 81377, Munich, Germany.
Strahlenther Onkol ; 194(7): 664-674, 2018 07.
Article em En | MEDLINE | ID: mdl-29523907
ABSTRACT

PURPOSE:

To evaluate the impact of shorter treatment times on intra-fractional motion of the prostate during external beam radiotherapy.

METHODS:

53 h of intra-fractional motion of the prostate were recorded in real-time by 4D ultrasound (4DUS) during 720 fractions in 28 patients, 14 of which whom treated with step-and-shoot intensity-modulated radiotherapy (IMRT) and 14 of whom were treated with volumetric arc therapy (VMAT).

RESULTS:

The average VMAT fraction was recorded for 2 min 43 s and was substantially shorter than the average step-and-shoot IMRT fraction at 6 min 13 s. Average radial displacement of the prostate per fraction was substantially and significantly reduced from 1.31 ± 1.28 mm (n = 357 step-and-shoot IMRT fractions) to 0.96 ± 1.04 mm (n = 363 VMAT fractions), p = 0.00004. Radial, vertical, and longitudinal root-mean-square (r. m. s.) error per fraction was reduced from 1.55 to 1.12 mm (-28%, p < 0.0001), from 1.16 to 0.77 mm (-34%, p < 0.0001), and from 0.79 to 0.56 mm (-29%, p = 0.0002), respectively. Lateral intra-fractional motion was generally small and did not differ significantly. The prostate remained during 95% of fraction time within 4.55 mm of the isocenter in case of step-and-shoot IMRT and within 2.45 mm in case of VMAT. The variance of displacements increased linearly with time, and the rate was the same for both step-and-shoot IMRT and VMAT patients.

CONCLUSIONS:

The position of the prostate changed less during shorter fractions, limiting fraction-average and end-of-fraction variance. This substantially and significantly reduced the impact of intra-fractional motion during shorter VMAT fractions as compared to longer step-and-shoot IMRT fractions.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias da Próstata / Adenocarcinoma / Radioterapia de Intensidade Modulada / Posicionamento do Paciente / Movimentos dos Órgãos Limite: Aged / Humans / Male Idioma: En Revista: Strahlenther Onkol Assunto da revista: NEOPLASIAS / RADIOTERAPIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias da Próstata / Adenocarcinoma / Radioterapia de Intensidade Modulada / Posicionamento do Paciente / Movimentos dos Órgãos Limite: Aged / Humans / Male Idioma: En Revista: Strahlenther Onkol Assunto da revista: NEOPLASIAS / RADIOTERAPIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha