Limitations of platform assays to measure serum 25OHD level impact on guidelines and practice decision making.
Rahme, Maya; Al-Shaar, Laila; Singh, Ravinder; Baddoura, Rafic; Halaby, Georges; Arabi, Asma; Habib, Robert H; Daher, Rose; Bassil, Darina; El-Ferkh, Karim; Hoteit, Maha; El-Hajj Fuleihan, Ghada.
Afiliação
  • Rahme M; Department of Internal Medicine, Division of Endocrinology, Calcium Metabolism and Osteoporosis Program, WHO Collaborating Center for Metabolic Bone Disorders, American University of Beirut Medical Center, Beirut, Lebanon.
  • Al-Shaar L; Department of Internal Medicine, Division of Endocrinology, Calcium Metabolism and Osteoporosis Program, WHO Collaborating Center for Metabolic Bone Disorders, American University of Beirut Medical Center, Beirut, Lebanon.
  • Singh R; Division of Clinical Biochemistry and Immunology, Mayo Clinic Laboratories, Mayo Clinic, Rochester, MN, USA.
  • Baddoura R; Department of Rheumatology, Hotel Dieu de France, St Joseph University, Beirut, Lebanon.
  • Halaby G; Department of Endocrinology, Hotel Dieu de France, St Joseph University, Beirut, Lebanon.
  • Arabi A; Department of Internal Medicine, Division of Endocrinology, Calcium Metabolism and Osteoporosis Program, WHO Collaborating Center for Metabolic Bone Disorders, American University of Beirut Medical Center, Beirut, Lebanon.
  • Habib RH; Department of Internal Medicine, American University of Beirut Medical Center, Beirut, Lebanon.
  • Daher R; Department of Pathology & Laboratory Medicine, American University of Beirut Medical Center, Beirut, Lebanon.
  • Bassil D; Department of Internal Medicine, Division of Endocrinology, Calcium Metabolism and Osteoporosis Program, WHO Collaborating Center for Metabolic Bone Disorders, American University of Beirut Medical Center, Beirut, Lebanon.
  • El-Ferkh K; Department of Internal Medicine, Division of Endocrinology, Calcium Metabolism and Osteoporosis Program, WHO Collaborating Center for Metabolic Bone Disorders, American University of Beirut Medical Center, Beirut, Lebanon.
  • Hoteit M; Department of Internal Medicine, Division of Endocrinology, Calcium Metabolism and Osteoporosis Program, WHO Collaborating Center for Metabolic Bone Disorders, American University of Beirut Medical Center, Beirut, Lebanon.
  • El-Hajj Fuleihan G; Department of Internal Medicine, Division of Endocrinology, Calcium Metabolism and Osteoporosis Program, WHO Collaborating Center for Metabolic Bone Disorders, American University of Beirut Medical Center, Beirut, Lebanon. Electronic address: gf01@aub.edu.lb.
Metabolism ; 89: 1-7, 2018 12.
Article em En | MEDLINE | ID: mdl-30227144
ABSTRACT
CONTEXT Liquid Chromatography Mass Spectroscopy (LC-MS/MS) is the preferred method to measure 25 hydroxyvitamin D (25OHD) levels, but laboratories are increasingly adopting automated platform assays.

OBJECTIVE:

We assessed the performance of commonly used automated immunoassays, with that of LC-MS/MS, and the National Institute of Standards and Technology (NIST) reference values, to measure 25OHD levels. METHODS/

SETTING:

We compared serum 25OHD levels obtained from 219 elderly subjects, enrolled in a vitamin D trial, using the Diasorin Liaison platform assay, and the tandem LC-MS/MS method. We also assessed the performance of the Diasorin and Roche automated assays, expressed as mean % bias from the NIST standards, based on the vitamin D External Quality Assessment Scheme (DEQAS) reports, from 2013 to 2017.

RESULTS:

Serum 25OHD levels were significantly lower in the Diasorin compared to LC-MS/MS assay at baseline, 18.5 ±â€¯7.8 vs 20.5 ±â€¯7.6 ng/ml (p < 0.001), and all other time points. Diasorin (25OHD) = 0.76 × LC-MS/MS (25OHD) + 4.3, R2 = 0.596. The absolute bias was independent of 25OHD values, and the pattern unfit for any cross-calibration. The proportion of subjects considered for vitamin D treatment based on pre-set cut-offs differed significantly between the 2 assays. There also was wide variability in the performance of both automated assays, compared to NIST reference values.

CONCLUSION:

The performance of most widely used automated assays is sub-optimal. Our findings underscore the pressing need to re-consider current practices with regard to 25OHD measurements, interpretation of results from research studies, meta-analyses, the development of vitamin D guidelines, and their relevance to optimizing health.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Vitamina D / Deficiência de Vitamina D / Tomada de Decisão Clínica Tipo de estudo: Guideline / Prognostic_studies Limite: Aged / Female / Humans / Male Idioma: En Revista: Metabolism Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Líbano

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Vitamina D / Deficiência de Vitamina D / Tomada de Decisão Clínica Tipo de estudo: Guideline / Prognostic_studies Limite: Aged / Female / Humans / Male Idioma: En Revista: Metabolism Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Líbano