Electrometer offset current due to scattered radiation.
Wegener, Sonja; Sauer, Otto A.
Afiliação
  • Wegener S; Department of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
  • Sauer OA; Department of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
J Appl Clin Med Phys ; 19(6): 274-281, 2018 Nov.
Article em En | MEDLINE | ID: mdl-30298980
ABSTRACT
Relative dose measurements with small ionization chambers in combination with an electrometer placed in the treatment room ("internal electrometer") show a large dependence on the polarity used. While this was observed previously for percent depth dose curves (PDDs), the effect has not been understood or preventable. To investigate the polarity dependence of internal electrometers used in conjunction with a small-volume ionization chamber, we placed an internal electrometer at a distance of 1 m from the isocenter and exposed it to different amounts of scattered radiation by varying the field size. We identified irradiation of the electrometer to cause a current of approximately -1 pA, regardless of the sign of the biasing voltage. For low-sensitivity detectors, such a current noticeably distorts relative dose measurements. To demonstrate how the current systematically changes PDDs, we collected measurements with nine ionization chambers of different volumes. As the chamber volume decreased, signal ratios at 20 and 10 cm depth (M20/M10) became smaller for positive bias voltage and larger for negative bias voltage. At the size of the iba CC04 (40 mm³) the difference of M20/M10 was around 1% and for the smallest studied chamber, the iba CC003 chamber (3 mm³), around 7% for a 10 × 10 cm² field. When the electrometer was moved further from the source or shielded, the additional current decreased. Consequently, PDDs at both polarities were brought into alignment at depth even for the 3 mm³ ionization chamber. The apparent polarity effect on PDDs and lateral beam profiles was reduced considerably by shielding the electrometer. Due to normalization the effect on output values was low. When measurements with a low-sensitivity probe are carried out in conjunction with an internal electrometer, we recommend careful monitoring of the particular setup by testing both polarities, and if deemed necessary, we suggest shielding the electrometer.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Radiometria / Espalhamento de Radiação / Eletrônica Limite: Humans Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Radiometria / Espalhamento de Radiação / Eletrônica Limite: Humans Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha