Test-retest repeatability and reproducibility of ADC measures by breast DWI: Results from the ACRIN 6698 trial.
J Magn Reson Imaging
; 49(6): 1617-1628, 2019 06.
Article
em En
| MEDLINE
| ID: mdl-30350329
ABSTRACT
BACKGROUND:
Quantitative diffusion-weighted imaging (DWI) MRI is a promising technique for cancer characterization and treatment monitoring. Knowledge of the reproducibility of DWI metrics in breast tumors is necessary to apply DWI as a clinical biomarker.PURPOSE:
To evaluate the repeatability and reproducibility of breast tumor apparent diffusion coefficient (ADC) in a multi-institution clinical trial setting, using standardized DWI protocols and quality assurance (QA) procedures. STUDY TYPE Prospective.SUBJECTS:
In all, 89 women from nine institutions undergoing neoadjuvant chemotherapy for invasive breast cancer. FIELD STRENGTH/SEQUENCE DWI was acquired before and after patient repositioning using a four b-value, single-shot echo-planar sequence at 1.5T or 3.0T. ASSESSMENT A QA procedure by trained operators assessed artifacts, fat suppression, and signal-to-noise ratio, and determine study analyzability. Mean tumor ADC was measured via manual segmentation of the multislice tumor region referencing DWI and contrast-enhanced images. Twenty cases were evaluated multiple times to assess intra- and interoperator variability. Segmentation similarity was assessed via the Sørenson-Dice similarity coefficient. STATISTICAL TESTS Repeatability and reproducibility were evaluated using within-subject coefficient of variation (wCV), intraclass correlation coefficient (ICC), agreement index (AI), and repeatability coefficient (RC). Correlations were measured by Pearson's correlation coefficients.RESULTS:
In all, 71 cases (80%) passed QA evaluation 44 at 1.5T, 27 at 3.0T; 60 pretreatment, 11 after 3 weeks of taxane-based treatment. ADC repeatability was excellent wCV = 4.8% (95% confidence interval [CI] 4.0, 5.7%), ICC = 0.97 (95% CI 0.95, 0.98), AI = 0.83 (95% CI 0.76, 0.87), and RC = 0.16 * 10-3 mm2 /sec (95% CI 0.13, 0.19). The results were similar across field strengths and timepoint subgroups. Reproducibility was excellent interreader ICC = 0.92 (95% CI 0.80, 0.97) and intrareader ICC = 0.91 (95% CI 0.78, 0.96). DATACONCLUSION:
Breast tumor ADC can be measured with excellent repeatability and reproducibility in a multi-institution setting using a standardized protocol and QA procedure. Improvements to DWI image quality could reduce loss of data in clinical trials. LEVEL OF EVIDENCE 2 Technical Efficacy Stage 1 J. Magn. Reson. Imaging 2019;491617-1628.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Mama
/
Neoplasias da Mama
/
Imagem de Difusão por Ressonância Magnética
/
Neoplasias
Tipo de estudo:
Clinical_trials
/
Guideline
/
Observational_studies
Limite:
Adult
/
Aged
/
Female
/
Humans
/
Middle aged
Idioma:
En
Revista:
J Magn Reson Imaging
Assunto da revista:
DIAGNOSTICO POR IMAGEM
Ano de publicação:
2019
Tipo de documento:
Article
País de afiliação:
Estados Unidos