Your browser doesn't support javascript.
loading
Accurate fatty acid composition estimation of adipose tissue in the abdomen based on bipolar multi-echo MRI.
Schneider, Manuel; Janas, Gemini; Lugauer, Felix; Hoppe, Elisabeth; Nickel, Dominik; Dale, Brian M; Kiefer, Berthold; Maier, Andreas; Bashir, Mustafa R.
Afiliação
  • Schneider M; Pattern Recognition Lab, Department of Computer Science, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.
  • Janas G; Radiology, Duke University Medical Center, Durham, North Carolina.
  • Lugauer F; Center for Advanced Magnetic Resonance Development, Duke University Medical Center, Durham, North Carolina.
  • Hoppe E; Pattern Recognition Lab, Department of Computer Science, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.
  • Nickel D; Pattern Recognition Lab, Department of Computer Science, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.
  • Dale BM; MR Applications Predevelopment, Siemens Healthcare GmbH, Erlangen, Germany.
  • Kiefer B; MR R&D Collaborations, Siemens Healthineers, Cary, North Carolina.
  • Maier A; MR Applications Predevelopment, Siemens Healthcare GmbH, Erlangen, Germany.
  • Bashir MR; Pattern Recognition Lab, Department of Computer Science, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.
Magn Reson Med ; 81(4): 2330-2346, 2019 04.
Article em En | MEDLINE | ID: mdl-30368904
ABSTRACT

PURPOSE:

To develop a bipolar multi-echo MRI method for the accurate estimation of the adipose tissue fatty acid composition (FAC) using clinically relevant protocols at clinical field strength.

METHODS:

The proposed technique jointly estimates confounding factors (field map, R2* , eddy-current phase) and triglyceride saturation state parameters by fitting multi-echo gradient echo acquisitions to a complex signal model. The noise propagation behavior was improved by applying a low-rank enforcing denoising technique and by addressing eddy-current-induced phase discrepancies analytically. The impact of the total echo train duration on the FAC parameter map accuracy was analyzed in an oil phantom at 3T. Accuracy and reproducibility assessment was based on in vitro oil phantom measurements at two field strengths (3T and 1.5T) and with two different protocols. Repeatability was assessed in vivo in patients (n = 8) with suspected fatty liver disease using test-retest acquisitions in the abdominal subcutaneous, perirenal and mesenteric fat depots.

RESULTS:

Echo train readout durations of at least five times the conventional in-phase time were required for accurate FAC estimation in areas of high fat content. In vitro, linear regression and Bland-Altman analyses demonstrated strong (r > 0.94) and significant (P â‰ª 0.01) correlations between measured and reference FACs for all acquisitions, with smaller overall intercepts and biases at 3T compared to 1.5T. In vivo, reported mean absolute differences between test and retest were 1.54%, 3.31%, and 2.63% for the saturated, mono-unsaturated, and poly-unsaturated fat component, respectively.

CONCLUSIONS:

Accurate and reproducible MRI-based FAC quantification within a breath-hold is possible at clinical field strengths.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Tecido Adiposo / Abdome / Ácidos Graxos Tipo de estudo: Guideline / Observational_studies / Prognostic_studies Limite: Adolescent / Adult / Child / Child, preschool / Female / Humans / Male / Middle aged Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Tecido Adiposo / Abdome / Ácidos Graxos Tipo de estudo: Guideline / Observational_studies / Prognostic_studies Limite: Adolescent / Adult / Child / Child, preschool / Female / Humans / Male / Middle aged Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Alemanha