Analytical performance of automated platelet counts and impact on platelet transfusion guidance in patients with acute leukemia.
Tantanate, Chaicharoen; Khowawisetsut, Ladawan; Sukapirom, Kasama; Pattanapanyasat, Kovit.
Afiliação
  • Tantanate C; a Department of Clinical Pathology, Faculty of Medicine Siriraj Hospital , Mahidol University , Bangkok , Thailand.
  • Khowawisetsut L; b Department of Parasitology, Faculty of Medicine Siriraj Hospital , Mahidol University , Bangkok , Thailand.
  • Sukapirom K; c Department of Research and Development , Siriraj Center of Research Excellence for Microparticle and Exosome in Diseases, Faculty of Medicine Siriraj Hospital, Mahidol University , Bangkok , Thailand.
  • Pattanapanyasat K; c Department of Research and Development , Siriraj Center of Research Excellence for Microparticle and Exosome in Diseases, Faculty of Medicine Siriraj Hospital, Mahidol University , Bangkok , Thailand.
Scand J Clin Lab Invest ; 79(3): 160-166, 2019 May.
Article em En | MEDLINE | ID: mdl-30761915
The aim of this study was to evaluate the performance of automated impedance platelet counts by Beckman Coulter LH780 (PLT-LH), Sysmex XN-3000 (PLT-XNi) and fluorescence method by Sysmex XN-3000 (PLT-F) in patients with acute leukemia. Blood specimens were subjected to platelet measurements by evaluated methods and then compared against the international reference method (IRM). Eighty-two blood specimens were included. Bland-Altman plots of the differences between the evaluated methods and IRM demonstrated mean biases of PLT-LH, PLT-XNi and PLT-F of 9 × 109/L, 11 × 109/L and 2 × 109/L, respectively. For platelet transfusion guidance, all evaluated methods had acceptable accuracy. For platelet transfusion guidance, the sensitivities of PLT-LH, PLT-XNi and PLT-F were 33.3, 25.0 and 83.3%, respectively, at a transfusion threshold of 10 × 109/L, and 73.1, 61.5 and 84.6%, respectively, at transfusion threshold of 20 × 109/L. High blast count was associated with inaccurate PLT-LH and PLT-XNi. In conclusion, the PLT-F demonstrated excellent performance for diagnosis of thrombocytopenia and for platelet transfusion guidance in the evaluated specimens from acute leukemia patients. With respect to clinical relevance, careful blood smear review is necessary in case of high blast counts.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Leucemia / Transfusão de Plaquetas Tipo de estudo: Guideline Limite: Adult / Female / Humans / Male Idioma: En Revista: Scand J Clin Lab Invest Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Tailândia

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Leucemia / Transfusão de Plaquetas Tipo de estudo: Guideline Limite: Adult / Female / Humans / Male Idioma: En Revista: Scand J Clin Lab Invest Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Tailândia