Assessment of imaging Cherenkov and scintillation signals in head and neck radiotherapy.
Alexander, Daniel A; Tendler, Irwin I; Bruza, Petr; Cao, Xu; Schaner, Philip E; Marshall, Bethany S; Jarvis, Lesley A; Gladstone, David J; Pogue, Brian W.
Afiliação
  • Alexander DA; Thayer School of Engineering, Dartmouth College, Hanover, NH, United States of America. Author to whom any correspondence should be addressed.
Phys Med Biol ; 64(14): 145021, 2019 07 18.
Article em En | MEDLINE | ID: mdl-31146269
ABSTRACT
The goal of this study was to test the utility of time-gated optical imaging of head and neck (HN) radiotherapy treatments to measure surface dosimetry in real-time and inform possible interfraction replanning decisions. The benefit of both Cherenkov and scintillator imaging in HN treatments is direct daily feedback on dose, with no change to the clinical workflow. Emission from treatment materials was characterized by measuring radioluminescence spectra during irradiation and comparing emission intensities relative to Cherenkov emission produced in phantoms and scintillation from small plastic targets. HN treatment plans were delivered to a phantom with bolus and mask present to measure impact on signal quality. Interfraction superficial tumor reduction was simulated on a HN phantom, and cumulative Cherenkov images were analyzed in the region of interest (ROI). HN human patient treatment was imaged through the mask and compared with the dose distribution calculated by the treatment planning system. The relative intensity of radioluminescence from the mask was found to be within 30% of the Cherenkov emission intensity from tissue-colored clay. A strong linear relationship between normalized cumulative Cherenkov intensity and tumor size was established ([Formula see text]). The presence of a mask above a scintillator ROI was found to decrease mean pixel intensity by >40% and increase distribution spread. Cherenkov imaging through mask material is shown to have potential for surface field verification and tracking of superficial anatomy changes between treatment fractions. Imaging of scintillating targets provides a direct imaging of surface dose on the patient and through transparent bolus material. The first imaging of a patient receiving HN radiotherapy was achieved with a signal map which qualitatively matches the surface dose plan.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Contagem de Cintilação / Algoritmos / Processamento de Imagem Assistida por Computador / Planejamento da Radioterapia Assistida por Computador / Imagens de Fantasmas / Imagem Óptica / Neoplasias de Cabeça e Pescoço Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Phys Med Biol Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Contagem de Cintilação / Algoritmos / Processamento de Imagem Assistida por Computador / Planejamento da Radioterapia Assistida por Computador / Imagens de Fantasmas / Imagem Óptica / Neoplasias de Cabeça e Pescoço Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Phys Med Biol Ano de publicação: 2019 Tipo de documento: Article