Your browser doesn't support javascript.
loading
Modular implant design affects metal ion release following metal-on-metal hip arthroplasty: a retrospective study on 75 cases.
La Barbera, L; D'apolito, R; Peretti, G M; Piergiovanni, M; Banfi, G; Zagra, L.
Afiliação
  • La Barbera L; Laboratory of Biological Structure Mechanics, Department of Chemistry, Materials and Chemical Engineering "Giulio Natta", Politecnico di Milano, Milan, Italy.
  • D'apolito R; IRCCS Istituto Ortopedico Galeazzi, Milan, Italy.
  • Peretti GM; IRCCS Istituto Ortopedico Galeazzi, Milan, Italy.
  • Piergiovanni M; Department of Biomedical Sciences for Health, University of Milan, Milan, Italy.
  • Banfi G; Laboratory of Biological Structure Mechanics, Department of Chemistry, Materials and Chemical Engineering "Giulio Natta", Politecnico di Milano, Milan, Italy.
  • Zagra L; IRCCS Istituto Ortopedico Galeazzi, Milan, Italy.
J Biol Regul Homeost Agents ; 33(2 Suppl. 1): 79-88. XIX Congresso Nazionale S.I.C.O.O.P. Societa' Italiana Chirurghi Ortopedici Dell'ospedalita' Privata Accreditata, 2019.
Article em En | MEDLINE | ID: mdl-31169007
ABSTRACT
Metal-on-Metal (MoM) total hip arthroplasty (THA) has been associated to wear and metal-ions release, controversially related to a variety of clinical complications. Little is known about the relevant design-dependent parameters involved in this process. The present study investigated the correlation between metal ion release in blood and revision rate as a function of (i) specific MoM implant modular design parameters, (i.e. acetabular cup and femoral head diameters, taper adapter material and size, femoral neck material and modularity and stem size); (ii) MoM bilaterality. Co and Cr ions concentration levels in blood of 75 patients were retrospectively-evaluated with a mean follow-up of 4.8 years (range 1.8-6.3). Patients were divided in a unilateral and a bilateral group. Statistical analysis was performed to find any significant difference related to acetabular cup diameter, femoral head diameter, taper adapter material/size, neck material/size and stem size. The bilateral MoM group had 4-times higher metal ion levels in blood than the unilateral one (p=0.017 only Cr), related to a higher revision rate (30% vs 20%) differences were 10-times higher particularly with a 48 mm femoral head diameter (p=0.012) and a Ti-alloy neck (p=0.041). Within the monolateral group using a shorter taper adapter and a shorter neutrally-oriented neck demonstrated higher ion levels (p=0.038 only Cr and p=0.008 only Co, respectively). The aforementioned design-features and MoM bilaterality are important risk-factors for metal-ion release in modular MoM THA.
Assuntos
Palavras-chave
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Artroplastia de Quadril / Próteses Articulares Metal-Metal / Prótese de Quadril Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: J Biol Regul Homeost Agents Assunto da revista: BIOLOGIA / BIOQUIMICA Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Itália
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Artroplastia de Quadril / Próteses Articulares Metal-Metal / Prótese de Quadril Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: J Biol Regul Homeost Agents Assunto da revista: BIOLOGIA / BIOQUIMICA Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Itália