Intermediate outcomes of transcatheter pulmonary valve replacement with the Edwards Sapien 3 valve - German experience.
Lehner, Anja; Dashkalova, Tsvetina; Ulrich, Sarah; Fernandez Rodriguez, Silvia; Mandilaras, Guido; Jakob, Andre; Dalla-Pozza, Robert; Fischer, Marcus; Schneider, Heike; Tarusinov, Gleb; Kampmann, Christoph; Hofbeck, Michael; Dähnert, Ingo; Kanaan, Majed; Haas, Nikolaus A.
Afiliação
  • Lehner A; Department for Pediatric Cardiology and Intensive Care, Medical Hospital of the University of Munich, LMU Ludwig Maximilians University Munich , Munich , Germany.
  • Dashkalova T; Department for Pediatric Cardiology and Intensive Care, Medical Hospital of the University of Munich, LMU Ludwig Maximilians University Munich , Munich , Germany.
  • Ulrich S; Department for Pediatric Cardiology and Intensive Care, Medical Hospital of the University of Munich, LMU Ludwig Maximilians University Munich , Munich , Germany.
  • Fernandez Rodriguez S; Department for Pediatric Cardiology and Intensive Care, Medical Hospital of the University of Munich, LMU Ludwig Maximilians University Munich , Munich , Germany.
  • Mandilaras G; Department for Pediatric Cardiology and Intensive Care, Medical Hospital of the University of Munich, LMU Ludwig Maximilians University Munich , Munich , Germany.
  • Jakob A; Department for Pediatric Cardiology and Intensive Care, Medical Hospital of the University of Munich, LMU Ludwig Maximilians University Munich , Munich , Germany.
  • Dalla-Pozza R; Department for Pediatric Cardiology and Intensive Care, Medical Hospital of the University of Munich, LMU Ludwig Maximilians University Munich , Munich , Germany.
  • Fischer M; Department for Pediatric Cardiology and Intensive Care, Medical Hospital of the University of Munich, LMU Ludwig Maximilians University Munich , Munich , Germany.
  • Schneider H; Department for Pediatric Cardiology and Intensive Care, Heart Center Goettingen, Georg-August-University Goettingen , Goettingen , Germany.
  • Tarusinov G; Department for Pediatric Cardiology, Heart Center Duisburg , Duisburg , Germany.
  • Kampmann C; Department for Pediatric Cardiology, University Medical Center Mainz , Mainz , Germany.
  • Hofbeck M; Department for Pediatric Cardiology, University Hospital Tuebingen , Tuebingen , Germany.
  • Dähnert I; Department for Pediatric Cardiology, Heart Center Leipzig , Leipzig , Germany.
  • Kanaan M; Center for Congenital Heart Defects, Heart and Diabetes Centre North Rhine Westphalia, Ruhr University Bochum , Bad Oeynhausen , Germany.
  • Haas NA; Department for Pediatric Cardiology and Intensive Care, Medical Hospital of the University of Munich, LMU Ludwig Maximilians University Munich , Munich , Germany.
Expert Rev Med Devices ; 16(9): 829-834, 2019 Sep.
Article em En | MEDLINE | ID: mdl-31432698
ABSTRACT

Background:

After encouraging results with the Edwards Sapien and XT valves, this study aimed to review procedural data and early outcomes for the Sapien 3 valves for transcatheter pulmonary valve replacement (TPVR).

Methods:

We performed a multicenter, retrospective analysis of cases who underwent a Sapien 3 TPVR between 2015 and 2017 in 7 centers in Germany with a follow-up of up to 2 years.

Results:

56 patients could be enrolled (weight 58,5 ± 25,0 kg; 53% Tetralogy of Fallot, 45% native RVOT). Most procedures were two-stage procedures (82,1%) with 100% prestenting. Valve sizes were 20 mm (n = 1), 23 mm (n = 15), 26 mm (n = 27), 29 mm (n = 13). Procedural success rate was 96.4%. Two patients underwent surgical valve implantation after balloon rupture during TPVR. Follow-up data were available up to 24-month post TPVR. The rate of patients with ? moderate and severe pulmonary regurgitation decreased to 0% after TPVR, peak systolic gradient decreased from 24,2 (SD±20,9) mmHg to 7,1 mmHg (SD±5,0). There were no endocarditis, severe tricuspid valve impairment or stent fractures.

Conclusions:

With the Edwards Sapien 3 valve, the patient pool for TPVR can be substantially extended. Continued data collection is necessary to verify long-term results.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Pulmonar / Cateterismo Cardíaco / Implante de Prótese de Valva Cardíaca Tipo de estudo: Observational_studies / Prognostic_studies Limite: Adolescent / Adult / Aged / Aged80 / Child / Child, preschool / Female / Humans / Male / Middle aged País/Região como assunto: Europa Idioma: En Revista: Expert Rev Med Devices Assunto da revista: DIAGNOSTICO POR IMAGEM / TERAPEUTICA Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Pulmonar / Cateterismo Cardíaco / Implante de Prótese de Valva Cardíaca Tipo de estudo: Observational_studies / Prognostic_studies Limite: Adolescent / Adult / Aged / Aged80 / Child / Child, preschool / Female / Humans / Male / Middle aged País/Região como assunto: Europa Idioma: En Revista: Expert Rev Med Devices Assunto da revista: DIAGNOSTICO POR IMAGEM / TERAPEUTICA Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Alemanha