Investigation of detector uniformity issues for Siemens Inspiration systems.
Baldelli, P; Keavey, E; Manley, M; Power, G; Phelan, N.
Afiliação
  • Baldelli P; BreastCheck, National Screening Service, Eccles Unit, 36 Eccles St., Dublin 7, Ireland. Electronic address: paola.baldelli@screeningservice.ie.
  • Keavey E; BreastCheck, National Screening Service, Western Unit, Newcastle Rd, Galway, Ireland.
  • Manley M; BreastCheck, National Screening Service, Southern Unit, Infirmary Rd, Ballintemple, Cork, Ireland.
  • Power G; BreastCheck, National Screening Service, Eccles Unit, 36 Eccles St., Dublin 7, Ireland.
  • Phelan N; BreastCheck, National Screening Service, Eccles Unit, 36 Eccles St., Dublin 7, Ireland.
Phys Med ; 69: 262-268, 2020 Jan.
Article em En | MEDLINE | ID: mdl-31927263
ABSTRACT

PURPOSE:

Detector uniformity is an important parameter in digital mammography to guarantee a level of image quality adequate for early detection of breast cancer. Many problems with digital systems have been determined through the uniformity measurement, primarily as a result of incorrect flat-field calibration and artifacts caused by image receptor defects. The European guidelines suggest a method for the image uniformity assessment based on measurement of Signal-to-Noise ratio (SNR) and Pixel Value (PV) across a uniform image. Nineteen mammography systems from the same manufacturer installed in our organization incorporate an a-Se direct conversion detector. Since their installation, instability and inconsistency of image uniformity has attracted medical physicist attention. A number of different tests have been carried out in order to understand and establish reasons for this instability.

METHODS:

Three different tests have been performed to evaluate the impact of the heel effect, detector temperature and ghosting on the uniformity images. All the tests are based on the acquisition of uniform images as suggested by the European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis.

RESULTS:

Results show that an increase in detector temperature produces an increase of SNR and decrease of uniformity. A further decrease of uniformity ranging between 20% and 30% is due to the ghosting while a decrease of about 10% is due the heel effect.

CONCLUSIONS:

X-ray tube, system geometry and detector have an impact on the system uniformity and an understanding of the contribution of each is necessary in order to obtain comparable image quality among all the systems.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias da Mama / Mamografia / Intensificação de Imagem Radiográfica Tipo de estudo: Guideline / Screening_studies Limite: Female / Humans Idioma: En Revista: Phys Med Assunto da revista: BIOFISICA / BIOLOGIA / MEDICINA Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias da Mama / Mamografia / Intensificação de Imagem Radiográfica Tipo de estudo: Guideline / Screening_studies Limite: Female / Humans Idioma: En Revista: Phys Med Assunto da revista: BIOFISICA / BIOLOGIA / MEDICINA Ano de publicação: 2020 Tipo de documento: Article