Your browser doesn't support javascript.
loading
How to choose a point-of-care testing for troponin.
Lin, Ya-Hui; Zhang, Yang; Liu, Yu-Tao; Cui, Kai; Kang, Jin-Suo; Zhou, Zhou.
Afiliação
  • Lin YH; Center of Laboratory Medicine, Beijing Key Laboratory for Molecular Diagnostics of Cardiovascular Diseases, State Key Laboratory of Cardiovascular Disease, National Center for Cardiovascular Diseases & Fuwai Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijin
  • Zhang Y; Center of Laboratory Medicine, Beijing Key Laboratory for Molecular Diagnostics of Cardiovascular Diseases, State Key Laboratory of Cardiovascular Disease, National Center for Cardiovascular Diseases & Fuwai Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijin
  • Liu YT; Center of Laboratory Medicine, Beijing Key Laboratory for Molecular Diagnostics of Cardiovascular Diseases, State Key Laboratory of Cardiovascular Disease, National Center for Cardiovascular Diseases & Fuwai Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijin
  • Cui K; Center of Laboratory Medicine, Beijing Key Laboratory for Molecular Diagnostics of Cardiovascular Diseases, State Key Laboratory of Cardiovascular Disease, National Center for Cardiovascular Diseases & Fuwai Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijin
  • Kang JS; Center of Laboratory Medicine, Beijing Key Laboratory for Molecular Diagnostics of Cardiovascular Diseases, State Key Laboratory of Cardiovascular Disease, National Center for Cardiovascular Diseases & Fuwai Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijin
  • Zhou Z; Center of Laboratory Medicine, Beijing Key Laboratory for Molecular Diagnostics of Cardiovascular Diseases, State Key Laboratory of Cardiovascular Disease, National Center for Cardiovascular Diseases & Fuwai Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijin
J Clin Lab Anal ; 34(7): e23263, 2020 Jul.
Article em En | MEDLINE | ID: mdl-32222055
ABSTRACT

BACKGROUND:

Point-of-care (POC) cTn assays are needed when the central laboratory is unable to provide timely results to the emergency department. Many POC devices are available. The prospect of choosing them is daunting. In order to provide a quick decision-making reference for POC cTn device selection comparing them to the central laboratory, seven POC devices commonly employed by emergency department were evaluated.

METHODS:

Firstly, we reviewed all devices package inserts. Secondly, we evaluated several POC cTn assays for imprecision, linearity, and correlation with central laboratory assays according to CLSI EP protocols. The linear regression analyses were performed only for the detectable concentrations. Five cTnI devices (Alere Triage, BioMerieux Vidas, Mitsubishi Pathfast, ReLIA TZ-301, and Radiometer AQT90) were evaluated against a contemporary cTnI assay (Beckman Access II Accu TnI). Two cTnT assays (Radiometer AQT90 and Roche Cobas h232) were compared to a high-sensitivity (hs) cTnT method (Roche Cobas e601).

RESULTS:

For cTn levels around the 99th percentile upper reference limits (URLs) of the comparator assays, imprecision could not be assessed for the Alere, BioMerieux, and Cobas h232 as they gave undetectable readings due to a lack of assay sensitivity. Imprecision (CV) was unacceptably high for the ReLIA (33.3%). On account of this precision metric, these four assays were deemed unsuitable. Regression analyses showed acceptable linearity for all the POC devices. The correlation coefficients for ReLIA, BioMerieux, Cobas h232, and Radiometer cTnT were >0.95. Unlike the cTnT devices, the cTnI assays employ different capture and detection antibodies leading to non-commutable results. The POC cTn results were concordant with their comparator-Radiometer cTnT 90%, Pathfast cTnI 85%, and Radiometer cTnI 75%.

CONCLUSION:

Our study provides the procedure and essential data to guide selection of a POC cTn device. Of the point-of-care devices, methods evaluated Radiometer AQT90 (cTnI and cTnT) and Pathfast might be considered.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Análise Química do Sangue / Troponina I / Troponina T / Testes Imediatos Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Revista: J Clin Lab Anal Assunto da revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Análise Química do Sangue / Troponina I / Troponina T / Testes Imediatos Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Revista: J Clin Lab Anal Assunto da revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2020 Tipo de documento: Article