Validation of thoracic aortic dimensions on ECG-triggered SSFP as alternative to contrast-enhanced MRA.
Snel, G J H; Hernandez, L M; Slart, R H J A; Nguyen, C T; Sosnovik, D E; van Deursen, V M; Dierckx, R A J O; Velthuis, B K; Borra, R J H; Prakken, N H J.
Afiliação
  • Snel GJH; Department of Radiology, University Medical Center Groningen, University of Groningen, Hanzeplein 1, 9713 GZ, Groningen, The Netherlands. g.j.h.snel@umcg.nl.
  • Hernandez LM; Department of Radiology, University Medical Center Groningen, University of Groningen, Hanzeplein 1, 9713 GZ, Groningen, The Netherlands.
  • Slart RHJA; Department of Nuclear Medicine and Molecular Imaging, University Medical Center Groningen, University of Groningen, Hanzeplein 1, 9713 GZ, Groningen, The Netherlands.
  • Nguyen CT; Department of Biomedical Photonic Imaging, University of Twente, Dienstweg 1, 7522 ND, Enschede, The Netherlands.
  • Sosnovik DE; Department of Radiology, Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Harvard Medical School, 149 13th Street, Charlestown, MA, 02129, USA.
  • van Deursen VM; Cardiovascular Research Center, Massachusetts General Hospital, Harvard Medical School, 149 13th Street, Charlestown, MA, 02129, USA.
  • Dierckx RAJO; Department of Radiology, Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Harvard Medical School, 149 13th Street, Charlestown, MA, 02129, USA.
  • Velthuis BK; Cardiovascular Research Center, Massachusetts General Hospital, Harvard Medical School, 149 13th Street, Charlestown, MA, 02129, USA.
  • Borra RJH; Division of Health Sciences and Technology, Harvard-MIT, 7 Massachusetts Avenue, Cambridge, MA, 02139, USA.
  • Prakken NHJ; Department of Cardiology, University Medical Center Groningen, University of Groningen, Hanzeplein 1, 9713 GZ, Groningen, The Netherlands.
Eur Radiol ; 30(11): 5794-5804, 2020 Nov.
Article em En | MEDLINE | ID: mdl-32506262
ABSTRACT

OBJECTIVES:

Assessment of thoracic aortic dimensions with non-ECG-triggered contrast-enhanced magnetic resonance angiography (CE-MRA) is accompanied with motion artefacts and requires gadolinium. To avoid both motion artefacts and gadolinium administration, we evaluated the similarity and reproducibility of dimensions measured on ECG-triggered, balanced steady-state free precession (SSFP) MRA as alternative to CE-MRA.

METHODS:

All patients, with varying medical conditions, referred for thoracic aortic examination between September 2016 and March 2018, who underwent non-ECG-triggered CE-MRA and SSFP-MRA (1.5 T) were retrospectively included (n = 30). Aortic dimensions were measured after double-oblique multiplanar reconstruction by two observers at nine landmarks predefined by literature guidelines. Image quality was scored at the sinus of Valsalva, mid-ascending aorta and mid-descending aorta by semi-automatically assessing the vessel sharpness.

RESULTS:

Aortic dimensions showed high agreement between non-ECG-triggered CE-MRA and SSFP-MRA (r = 0.99, p < 0.05) without overestimation or underestimation of aortic dimensions in SSFP-MRA (mean difference, 0.1 mm; limits of agreement, - 1.9 mm and 1.9 mm). Intra- and inter-observer variabilities were significantly smaller with SSFP-MRA for the sinus of Valsalva and sinotubular junction. Image quality of the sinus of Valsalva was significantly better with SSFP-MRA, as fewer images were of impaired quality (3/30) than in CE-MRA (21/30). Reproducibility of dimensions was significantly better in images scored as good quality compared to impaired quality in both sequences.

CONCLUSIONS:

Thoracic aortic dimensions measured on SSFP-MRA and non-ECG-triggered CE-MRA were similar. As expected, SSFP-MRA showed better reproducibility close to the aortic root because of lesser motion artefacts, making it a feasible non-contrast imaging alternative. KEY POINTS • SSFP-MRA provides similar dimensions as non-ECG-triggered CE-MRA. • Intra- and inter-observer reproducibilities improve for the sinus of Valsalva and sinotubular junction with SSFP-MRA. • ECG-triggered SSFP-MRA shows better image quality for landmarks close to the aortic root in the absence of cardiac motion.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aorta Torácica / Artefatos / Angiografia por Ressonância Magnética / Imageamento Tridimensional / Eletrocardiografia Tipo de estudo: Observational_studies Limite: Adult / Female / Humans / Male Idioma: En Revista: Eur Radiol Assunto da revista: RADIOLOGIA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aorta Torácica / Artefatos / Angiografia por Ressonância Magnética / Imageamento Tridimensional / Eletrocardiografia Tipo de estudo: Observational_studies Limite: Adult / Female / Humans / Male Idioma: En Revista: Eur Radiol Assunto da revista: RADIOLOGIA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Holanda