Your browser doesn't support javascript.
loading
The Application of Digital Volume Correlation (DVC) to Evaluate Strain Predictions Generated by Finite Element Models of the Osteoarthritic Humeral Head.
Kusins, Jonathan; Knowles, Nikolas; Columbus, Melanie; Oliviero, Sara; Dall'Ara, Enrico; Athwal, George S; Ferreira, Louis M.
Afiliação
  • Kusins J; Department of Mechanical and Materials Engineering, Western University, London, ON, Canada.
  • Knowles N; Roth|McFarlane Hand and Upper Limb Centre, St. Joseph's Health Care, London, ON, Canada.
  • Columbus M; Department of Radiology, Cumming School of Medicine, University of Calgary, Calgary, AB, Canada.
  • Oliviero S; University of Calgary, Calgary, AB, Canada.
  • Dall'Ara E; Department of Oncology and Metabolism and INSIGNEO Institute for In Silico Medicine, University of Sheffield, Sheffield, UK.
  • Athwal GS; Department of Oncology and Metabolism and INSIGNEO Institute for In Silico Medicine, University of Sheffield, Sheffield, UK.
  • Ferreira LM; Roth|McFarlane Hand and Upper Limb Centre, St. Joseph's Health Care, London, ON, Canada.
Ann Biomed Eng ; 48(12): 2859-2869, 2020 Dec.
Article em En | MEDLINE | ID: mdl-32572730
Continuum-level finite element models (FEMs) of the humerus offer the ability to evaluate joint replacement designs preclinically; however, experimental validation of these models is critical to ensure accuracy. The objective of the current study was to quantify experimental full-field strain magnitudes within osteoarthritic (OA) humeral heads by combining mechanical loading with volumetric microCT imaging and digital volume correlation (DVC). The experimental data was used to evaluate the accuracy of corresponding FEMs. Six OA humeral head osteotomies were harvested from patients being treated with total shoulder arthroplasty and mechanical testing was performed within a microCT scanner. MicroCT images (33.5 µm isotropic voxels) were obtained in a pre- and post-loaded state and BoneDVC was used to quantify full-field experimental strains (≈ 1 mm nodal spacing, accuracy = 351 µstrain, precision = 518 µstrain). Continuum-level FEMs with two types of boundary conditions (BCs) were simulated: DVC-driven and force-driven. Accuracy of the FEMs was found to be sensitive to the BC simulated with better agreement found with the use of DVC-driven BCs (slope = 0.83, r2 = 0.80) compared to force-driven BCs (slope = 0.22, r2 = 0.12). This study quantified mechanical strain distributions within OA trabecular bone and demonstrated the importance of BCs to ensure the accuracy of predictions generated by corresponding FEMs.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Osteoartrite / Articulação do Ombro / Cabeça do Úmero / Osso Esponjoso Tipo de estudo: Prognostic_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Ann Biomed Eng Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Canadá

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Osteoartrite / Articulação do Ombro / Cabeça do Úmero / Osso Esponjoso Tipo de estudo: Prognostic_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Ann Biomed Eng Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Canadá