The impact of scanning data measurements on the Acuros dose calculation algorithm configuration.
Radiat Oncol
; 15(1): 169, 2020 Jul 10.
Article
em En
| MEDLINE
| ID: mdl-32650815
BACKGROUND: Many dose calculation algorithms for radiotherapy planning need to be configured for each clinical beam using pre-defined measurements. An optimization process adjusts the physical parameters able to estimate the energy released in the medium in any geometrical condition. This work investigates the impact of measured input data quality on the configuration of the type "c" Acuros-XB dose calculation algorithm in the Eclipse (Varian Medical Systems) treatment planning system. METHODS: Different datasets were acquired with the BeamScan water phantom (PTW) to configure 6 MV beams, for both flattened (6X) and flattening filter free mode (6FFF) for a Varian TrueBeam: (i) a correct dataset measured using a Semiflex-3D ion chamber, (ii) a set in missing lateral scatter conditions (MLS), (iii) a set with incorrect effective point of measurement (EPoM), (iv) sets acquired with PinPoint-3D chamber, DiodeP, microDiamond detectors. The Acuros-XB dose calculation algorithm (version 15.6) was configured using the reference dataset, the sets measured with the different detectors, with intentional errors, and using the representative beam data (RBD) made available by the vendor. The physical parameters obtained from each optimization process (spectrum, mean radial energy, electron contamination), were analyzed and compared. Calculated data were finally compared against the input and reference measurements. RESULTS: Concerning the physical parameters, the configurations presenting the largest differences were the MLS conditions (mean radial energy) and the incorrect EPoM (electron contamination). The calculation doses relative to the input data present low accuracy, with mean differences > 2% in some conditions. The PinPoint-3D ion chamber presented lower accuracy for the 6FFF beam. Regarding the RBD, calculations compared well with the input data used for the configuration, but not with the reference data. CONCLUSION: The MLS conditions and the incorrect setting of the EPoM lead to erroneous configurations and should be avoided. The choice of an appropriate detector is important. Whenever the representative beam data is used, a careful check under more clinical geometrical conditions is advised.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Planejamento da Radioterapia Assistida por Computador
/
Neoplasias
Tipo de estudo:
Prognostic_studies
Limite:
Humans
Idioma:
En
Revista:
Radiat Oncol
Assunto da revista:
NEOPLASIAS
/
RADIOTERAPIA
Ano de publicação:
2020
Tipo de documento:
Article
País de afiliação:
Itália