Radiation Exposure of Surgical Team During Endourological Procedures: International Atomic Energy Agency-South-Eastern European Group for Urolithiasis Research Study.
J Endourol
; 35(5): 574-582, 2021 05.
Article
em En
| MEDLINE
| ID: mdl-32791856
ABSTRACT
Introduction:
Fluoroscopy-guided endourology procedures require proper radiation protection to minimize radiation risk. This multicenter study aimed at investigating radiation protection practice and related radiation exposure of operating team members. Materials andMethods:
Six endourology centers from the South-Eastern European Group for Urolithiasis Research answered questionnaires and collected data of 315 procedures performed within a 3-months period, with simultaneous measurement of dose to staff and dose area product (DAP) to patient. A pair of calibrated personal dosimeters, one for body and one for eye-lens dose, was worn by all key staff members. Dosimeters were centrally calibrated, measured, and analyzed.Results:
The annual workload ranged from 173 to 865 procedures per center. Practice of personal dose monitoring and use of radiation protection shielding was found to be inconsistent. Lead aprons and thyroid collars were used by all, whereas protective eyewear was used in only half of centers. Due to the regular use of protective aprons, the whole-body dose of all 44 monitored staff members was safely below the regulatory dose limits. Eye-lens dose of 17 (14 urologists and 3 assisting staff) was above the dosimeter detection level, and dose per procedure varied from <10 to 63 µSv. The highest annual eye-lens dose of 13.5 mSv was found for the surgeon in the busiest department by using an over-the-couch X-ray tube without a ceiling suspended screen. Working closer to patient body with no protection resulted in a six-time higher eye-lens dose per DAP for a surgeon compared with others in the same center. Lower eye-dose per procedure was associated with lower DAP to patient and with the use of an under-the-couch tube, lower fluoroscopy pulse rate, collimation, fluoroscopy time, and acquired images.Conclusions:
The study results call for the need to establish standard protocols about use of fluoroscopy during endourology procedures and to increase radiation protection knowledge and awareness of surgical staff.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Exposição Ocupacional
/
Exposição à Radiação
/
Urolitíase
/
Energia Nuclear
Tipo de estudo:
Clinical_trials
/
Guideline
/
Qualitative_research
Limite:
Humans
Idioma:
En
Revista:
J Endourol
Assunto da revista:
UROLOGIA
Ano de publicação:
2021
Tipo de documento:
Article
País de afiliação:
Áustria