Your browser doesn't support javascript.
loading
Security millimetre wave body scanner safe for patients with leadless pacemakers or subcutaneous implantable cardioverter-defibrillators.
Blazek, Patrick; O'Connor, Matthew J; Weigand, Severin; Grebmer, Christian; Reif, Sebastian; Fichtner, Stephanie; Friedrich, Lena; Reents, Tilko; Kottmaier, Marc; Semmler, Verena; Bourier, Felix; Schunkert, Heribert; Deisenhofer, Isabel; Kolb, Christof; Lennerz, Carsten.
Afiliação
  • Blazek P; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Klinik an der Technischen Universität München, Munich, Germany.
  • O'Connor MJ; Department of Cardiology, Wellington Hospital, Wellington, New Zealand.
  • Weigand S; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Klinik an der Technischen Universität München, Munich, Germany.
  • Grebmer C; DZHK (German Centre for Cardiovascular Research), Partner Site Munich Heart Alliance, Munich, Germany.
  • Reif S; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Klinik an der Technischen Universität München, Munich, Germany.
  • Fichtner S; Klinik für Kardiologie und Internistische Intensivmedizin, Klinikum Bogenhausen, Munich, Germany.
  • Friedrich L; Klinikum Großhadern, Medizinische Klinik und Poliklinik I, Klinikum der Universität München, Munich, Germany.
  • Reents T; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Klinik an der Technischen Universität München, Munich, Germany.
  • Kottmaier M; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Klinik an der Technischen Universität München, Munich, Germany.
  • Semmler V; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Klinik an der Technischen Universität München, Munich, Germany.
  • Bourier F; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Klinik an der Technischen Universität München, Munich, Germany.
  • Schunkert H; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Klinik an der Technischen Universität München, Munich, Germany.
  • Deisenhofer I; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Klinik an der Technischen Universität München, Munich, Germany.
  • Kolb C; DZHK (German Centre for Cardiovascular Research), Partner Site Munich Heart Alliance, Munich, Germany.
  • Lennerz C; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Klinik an der Technischen Universität München, Munich, Germany.
J Interv Card Electrophysiol ; 61(3): 603-607, 2021 Sep.
Article em En | MEDLINE | ID: mdl-32860605
ABSTRACT

PURPOSE:

This study was designed to evaluate the electromagnetic interference (EMI) effects and safety of the new security screening millimetre wave body scanners (MWBSs) for patients with rare cardiac implantable electronic devices (CIEDs).

METHODS:

We identified 73 patients with either entirely subcutaneous implantable cardioverter-defibrillators (S-ICD) or leadless pacemakers (LPM) attending routine device follow-up. CIED programming was optimised for the detection of EMI occurrence, and high-voltage therapy was disabled. Patients then underwent millimetre wave body scans under continuous ECG monitoring. Scanning was performed at the recommended distance as well as in close proximity to the scanner emulating accidental exposure. CIED function was observed for EMI effects.

RESULTS:

There were no episodes of inhibition of pacing in the leadless pacemaker subgroup, no oversensing in the S-ICD subgroup and no spontaneous device reprogramming in any group. There was no change in pacing or sensing thresholds, and S-ICD vector eligibility remained unchanged after scanning with the MWBS. No CIEDs were identified by the MWBS during the study.

CONCLUSION:

No EMI events were detected during the use of MWBSs by patients with either S-ICDs or LPMs. This data should be reassuring for patients suggesting that they can undergo security body scans without worries or disclosure of their CIED status.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Desfibriladores Implantáveis Limite: Humans Idioma: En Revista: J Interv Card Electrophysiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Desfibriladores Implantáveis Limite: Humans Idioma: En Revista: J Interv Card Electrophysiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Alemanha