Development of a Sensitive High-Resolution Mass Spectrometry Approach for Urea Nitrogen Quantitation in Small Volumes of Bronchoalveolar Lavage Fluid (BALF).
Ulmer, Candice Z; Smith, Bianca; Thonkulpitak, Janet; Hardin, Joshua; Danilenko, Uliana; Frame, Tunde; Cheng, Po-Yung; Vesper, Hubert W.
Afiliação
  • Ulmer CZ; Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, 4770 Buford Hwy NE, MS F25, Atlanta, Georgia 30341, United States.
  • Smith B; Battelle, 505 King Avenue, Columbus, Ohio 43201, United States.
  • Thonkulpitak J; Oak Ridge Institute for Science and Education, P.O. Box 117, Oak Ridge, Tennessee 37831, United States.
  • Hardin J; Oak Ridge Institute for Science and Education, P.O. Box 117, Oak Ridge, Tennessee 37831, United States.
  • Danilenko U; Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, 4770 Buford Hwy NE, MS F25, Atlanta, Georgia 30341, United States.
  • Frame T; Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, 4770 Buford Hwy NE, MS F25, Atlanta, Georgia 30341, United States.
  • Cheng PY; Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, 4770 Buford Hwy NE, MS F25, Atlanta, Georgia 30341, United States.
  • Vesper HW; Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, 4770 Buford Hwy NE, MS F25, Atlanta, Georgia 30341, United States.
J Am Soc Mass Spectrom ; 31(11): 2270-2276, 2020 Nov 04.
Article em En | MEDLINE | ID: mdl-32931276
ABSTRACT
A sensitive, selective, and quantitative method incorporating high-resolution mass spectrometry was developed for the determination of blood urea nitrogen (BUN) in bronchoalveolar lavage fluid. The method requires no sample cleanup or derivatization prior to analysis. High-performance liquid chromatography (HPLC) on a Hypersil Gold PFP column (100 × 3 mm, 3 µm particle size) connected to a C18 guard column was employed for a 10 min chromatographic separation. The detection of urea was achieved using a Thermo Scientific Q-Exactive Plus instrument incorporating selected ion monitoring (SIM) modes for the protonated adduct of urea. The urea analytical measuring range for the method is 0.047-17.134 mg/dL, resulting in a BUN analytical measurement range of 0.022-8.007 mg/dL, which allows for quantitation over 3 orders of magnitude (R2 = 0.999). In addition, the method is suitable for small sample volumes (15 µL) with a high level of accuracy, precision, and specificity.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espectrometria de Massas / Ureia / Líquido da Lavagem Broncoalveolar / Cromatografia Líquida de Alta Pressão / Nitrogênio Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: J Am Soc Mass Spectrom Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espectrometria de Massas / Ureia / Líquido da Lavagem Broncoalveolar / Cromatografia Líquida de Alta Pressão / Nitrogênio Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: J Am Soc Mass Spectrom Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Estados Unidos