Influence of gender on the performance of urine dipstick and automated urinalysis in the diagnosis of urinary tract infections at the emergency department.
Middelkoop, S J M; van Pelt, L J; Kampinga, G A; Ter Maaten, J C; Stegeman, C A.
Afiliação
  • Middelkoop SJM; University of Groningen, University Medical Center Groningen, Department of Internal Medicine, Division of Nephrology, Groningen, the Netherlands. Electronic address: s.j.m.middelkoop@umcg.nl.
  • van Pelt LJ; University of Groningen, University Medical Center Groningen, Department of Laboratory Medicine, Groningen, the Netherlands.
  • Kampinga GA; University of Groningen, University Medical Center Groningen, Department of Medical Microbiology and Infection Prevention, Groningen, the Netherlands.
  • Ter Maaten JC; University of Groningen, University Medical Center Groningen, Department of Internal Medicine, Division of Emergency Medicine, Groningen, the Netherlands.
  • Stegeman CA; University of Groningen, University Medical Center Groningen, Department of Internal Medicine, Division of Nephrology, Groningen, the Netherlands.
Eur J Intern Med ; 87: 44-50, 2021 May.
Article em En | MEDLINE | ID: mdl-33775508
ABSTRACT

BACKGROUND:

Urinary tract infections (UTIs) are frequently encountered at the Emergency Department (ED). Given the anatomical differences between men and women, we aimed to clarify differences in the diagnostic performance of urinary parameters at the ED.

METHODS:

A cohort study of adults presenting at the ED with fever and/or clinical suspected UTI. Performance of urine dipstick (UD) and automated urinalysis (UF-1000i) were analysed for the total study population and men and women separately. We focused on 1) UTI diagnosis and 2) positive urine culture (UC, ≥105 CFU/ml) as outcome.

RESULTS:

In 360 of 917 cases (39.3%) UTI was established (men/women 35.1%/43.6%). Diagnostic accuracy of UD was around 10% lower in women compared to men. Median automated leucocyte and bacterial count were higher in women compared to men. Diagnostic performance by receiver operating analysis was 0.851 for leucocytes (men/women 0.879/0.817) and 0.850 for bacteria (men/women 0.898/0.791). At 90% sensitivity, cut-off values of leucocyte count (men 60/µL, women 43/µL), and bacterial count (men 75/µL, women 139/µL) showed performance differences in favour of men. In both men and women, diagnostic performance using specified cut-off values was not different between normal and non-normal bladder evacuation. UC was positive in 327 cases (men/women 149/178), as with UTI diagnosis, diagnostic values in men outperformed women.

CONCLUSIONS:

Overall diagnostic accuracy of urinary parameters for diagnosing UTI is higher in men. The described differences in cut-off values for leukocyte and bacterial counts for diagnosing UTI necessitates gender-specific cut-off values, probably reflecting the influence of anatomical and urogenital differences.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Infecções Urinárias / Urinálise Tipo de estudo: Diagnostic_studies / Etiology_studies / Incidence_studies / Observational_studies / Risk_factors_studies Limite: Adult / Female / Humans / Male Idioma: En Revista: Eur J Intern Med Assunto da revista: MEDICINA INTERNA Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Infecções Urinárias / Urinálise Tipo de estudo: Diagnostic_studies / Etiology_studies / Incidence_studies / Observational_studies / Risk_factors_studies Limite: Adult / Female / Humans / Male Idioma: En Revista: Eur J Intern Med Assunto da revista: MEDICINA INTERNA Ano de publicação: 2021 Tipo de documento: Article