Establishment and evaluation of a rat model of extracorporeal membrane oxygenation (ECMO) thrombosis using a 3D-printed mock-oxygenator.
Umei, Nao; Lai, Angela; Miller, Jennifer; Shin, Suji; Roberts, Kalliope; Ai Qatarneh, Saif; Ichiba, Shingo; Sakamoto, Atsuhiro; Cook, Keith E.
Afiliação
  • Umei N; Department of Anesthesiology and Pain Medicine, Graduate School of Medicine, Nippon Medical School, 1-1-5, Sendagi, Bunkyo-ku, Tokyo, 113-8603, Japan. u-nao@nms.ac.jp.
  • Lai A; Department of Anesthesiology, Nippon Medical School, 1-1-5, Sendagi, Bunkyo-ku, Tokyo, 113-8603, Japan. u-nao@nms.ac.jp.
  • Miller J; Department of Surgical Intensive Care Medicine, Nippon Medical School, 1-1-5, Sendagi, Bunkyo-ku, Tokyo, 113-8603, Japan. u-nao@nms.ac.jp.
  • Shin S; Department of Biomedical Engineering, Carnegie Mellon University, 5000 Forbes Avenue, Pittsburgh, PA, 15213, USA. u-nao@nms.ac.jp.
  • Roberts K; Department of Biomedical Engineering, Carnegie Mellon University, 5000 Forbes Avenue, Pittsburgh, PA, 15213, USA.
  • Ai Qatarneh S; Department of Biomedical Engineering, Carnegie Mellon University, 5000 Forbes Avenue, Pittsburgh, PA, 15213, USA.
  • Ichiba S; Department of Biomedical Engineering, Carnegie Mellon University, 5000 Forbes Avenue, Pittsburgh, PA, 15213, USA.
  • Sakamoto A; Department of Biomedical Engineering, Carnegie Mellon University, 5000 Forbes Avenue, Pittsburgh, PA, 15213, USA.
  • Cook KE; Department of Biomedical Engineering, Carnegie Mellon University, 5000 Forbes Avenue, Pittsburgh, PA, 15213, USA.
J Transl Med ; 19(1): 179, 2021 04 28.
Article em En | MEDLINE | ID: mdl-33910585
ABSTRACT

BACKGROUND:

Extracorporeal membrane oxygenation (ECMO) research using large animals requires a significant amount of resources, slowing down the development of new means of ECMO anticoagulation. Therefore, this study developed and evaluated a new rat ECMO model using a 3D-printed mock-oxygenator.

METHODS:

The circuit consisted of tubing, a 3D-printed mock-oxygenator, and a roller pump. The mock-oxygenator was designed to simulate the geometry and blood flow patterns of the fiber bundle in full-scale oxygenators but with a low (2.5 mL) priming volume. Rats were placed on arteriovenous ECMO at a 1.9 mL/min flow rate at two different heparin doses (n = 3 each) low (15 IU/kg/h for eight hours) versus high (50 IU/kg/h for one hour followed by 25 IU/kg/h for seven hours). The experiment continued for eight hours or until the mock-oxygenator failed. The mock-oxygenator was considered to have failed when its blood flow resistance reached three times its baseline resistance.

RESULTS:

During ECMO, rats maintained near-normal mean arterial pressure and arterial blood gases with minimal hemodilution. The mock-oxygenator thrombus weight was significantly different (p < 0.05) between the low (0.02 ± 0.006 g) and high (0.003 ± 0.001 g) heparin delivery groups, and blood flow resistance was also larger in the low anticoagulation group.

CONCLUSIONS:

This model is a simple, inexpensive system for investigating new anticoagulation agents for ECMO and provides low and high levels of anticoagulation that can serve as control groups for future studies.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Trombose / Oxigenação por Membrana Extracorpórea Limite: Animals Idioma: En Revista: J Transl Med Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Trombose / Oxigenação por Membrana Extracorpórea Limite: Animals Idioma: En Revista: J Transl Med Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Japão