COVID-19 in the Clinic: Aerosol Containment Mask for Endoscopic Otolaryngologic Clinic Procedures.
Ference, Elisabeth H; Kim, Wihan; Oghalai, John S; Kim, Jee-Hong; Applegate, Brian E.
Afiliação
  • Ference EH; Caruso Department of Otolaryngology-Head and Neck Surgery, Keck School of Medicine of University of Southern California, Los Angeles, California, USA.
  • Kim W; Caruso Department of Otolaryngology-Head and Neck Surgery, Keck School of Medicine of University of Southern California, Los Angeles, California, USA.
  • Oghalai JS; Caruso Department of Otolaryngology-Head and Neck Surgery, Keck School of Medicine of University of Southern California, Los Angeles, California, USA.
  • Kim JH; Caruso Department of Otolaryngology-Head and Neck Surgery, Keck School of Medicine of University of Southern California, Los Angeles, California, USA.
  • Applegate BE; Caruso Department of Otolaryngology-Head and Neck Surgery, Keck School of Medicine of University of Southern California, Los Angeles, California, USA.
Otolaryngol Head Neck Surg ; 166(5): 850-857, 2022 05.
Article em En | MEDLINE | ID: mdl-34154484
ABSTRACT

OBJECTIVE:

To create an aerosol containment mask (ACM) that contains aerosols during common otolaryngologic endoscopic procedures while protecting patients from environmental aerosols. STUDY

DESIGN:

Bench testing.

SETTING:

Mannequin testing.

METHODS:

The mask was designed in SolidWorks and 3-dimensional printed. Mannequins were fitted with a nebulizer to generate aerosols. Commercial particle counters were used to measure mask performance.

RESULTS:

The ACM has 2 ports on either side for instruments and endoscopes, a port for a filter, and a port that can evacuate aerosols contained within the mask via a standard suction pump. The mask contained aerosols on a mannequin with and without facial hair when the suction was set to 18.5 L/min. Other types of masks demonstrated substantial aerosol leakage under similar conditions. In a subsequent experiment, the ACM contained aerosols generated by a nebulizer up to the saturation of the particle detector without measurable leakage with or without suction.

CONCLUSION:

The ACM will accommodate rigid and flexible endoscopes plus instruments and prevent leakage of patient-generated aerosols, thus avoiding contamination of the room and protecting health care workers from airborne contagions. LEVEL OF EVIDENCE 2.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Otolaringologia / COVID-19 Limite: Humans Idioma: En Revista: Otolaryngol Head Neck Surg Assunto da revista: OTORRINOLARINGOLOGIA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Otolaringologia / COVID-19 Limite: Humans Idioma: En Revista: Otolaryngol Head Neck Surg Assunto da revista: OTORRINOLARINGOLOGIA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Estados Unidos