Manifestation of aberrations in full-field optical coherence tomography.
Opt Express ; 29(14): 22044-22065, 2021 Jul 05.
Article em En | MEDLINE | ID: mdl-34265978
We report on a theoretical model for image formation in full-field optical coherence tomography (FFOCT). Because the spatial incoherence of the illumination acts as a virtual confocal pinhole in FFOCT, its imaging performance is equivalent to a scanning time-gated coherent confocal microscope. In agreement with optical experiments enabling a precise control of aberrations, FFOCT is shown to have nearly twice the resolution of standard imaging at moderate aberration level. Beyond a rigorous study on the sensitivity of FFOCT with respect to aberrations, this theoretical model paves the way towards an optimized design of adaptive optics and computational tools for high-resolution and deep imaging of biological tissues.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia de Coerência Óptica / Óptica e Fotônica / Modelos Teóricos Limite: Humans Idioma: En Revista: Opt Express Assunto da revista: OFTALMOLOGIA Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia de Coerência Óptica / Óptica e Fotônica / Modelos Teóricos Limite: Humans Idioma: En Revista: Opt Express Assunto da revista: OFTALMOLOGIA Ano de publicação: 2021 Tipo de documento: Article