Balloon-Expandable TAVR Prosthesis Dislocates Into the Ascending Aorta.
Vendrik, Jeroen; van den Boogert, Thomas P W; Koch, Karel T; Baan, Jan.
Afiliação
  • Vendrik J; Heart Centre (part of the Amsterdam Cardiovascular Sciences), Amsterdam University Medical Centre, University of Amsterdam, Amsterdam, the Netherlands.
  • van den Boogert TPW; Heart Centre (part of the Amsterdam Cardiovascular Sciences), Amsterdam University Medical Centre, University of Amsterdam, Amsterdam, the Netherlands.
  • Koch KT; Heart Centre (part of the Amsterdam Cardiovascular Sciences), Amsterdam University Medical Centre, University of Amsterdam, Amsterdam, the Netherlands.
  • Baan J; Heart Centre (part of the Amsterdam Cardiovascular Sciences), Amsterdam University Medical Centre, University of Amsterdam, Amsterdam, the Netherlands.
JACC Case Rep ; 1(2): 101-104, 2019 Aug.
Article em En | MEDLINE | ID: mdl-34316759
ABSTRACT
This case report underlines the complexity of the transcatheter aortic valve replacement (TAVR) procedure where rare complications sometimes are inevitable, even in experienced hands. Supra-annular dislocation of the balloon-expandable prosthesis was caused by loss of capture of the temporary transvenous pacemaker lead and treated successfully by retracting it towards the abdominal aorta. (Level of Difficulty Advanced.).