Evaluation on Effectiveness of a New System as well as Analysis on Optimal Horizontal Eye Position for Vertical Video Head Impulse Test.
Jiang, Tao; Li, Fei; Yu, Jing; Huang, Ruo-Nan; Gao, Rui; Shang, Chen-Yong; Zhuang, Jian-Hua; Li, Wen-Yan.
Afiliação
  • Jiang T; ENT Institute and Otorhinolaryngology Department of Eye & ENT Hospital, State Key Laboratory of Medical Neurobiology and MOE Frontiers Center for Brain Science, Fudan University, Shanghai, 200031, China.
  • Li F; NHC Key Laboratory of Hearing Medicine (Fudan University), Shanghai, 200031, China.
  • Yu J; Department of Neurology, Changzheng Hospital, Naval Medical University, Shanghai, 200003, China.
  • Huang RN; ENT Institute and Otorhinolaryngology Department of Eye & ENT Hospital, State Key Laboratory of Medical Neurobiology and MOE Frontiers Center for Brain Science, Fudan University, Shanghai, 200031, China.
  • Gao R; ENT Institute and Otorhinolaryngology Department of Eye & ENT Hospital, State Key Laboratory of Medical Neurobiology and MOE Frontiers Center for Brain Science, Fudan University, Shanghai, 200031, China.
  • Shang CY; Department of Neurology, Changzheng Hospital, Naval Medical University, Shanghai, 200003, China.
  • Zhuang JH; Department of Neurology, Changzheng Hospital, Naval Medical University, Shanghai, 200003, China.
  • Li WY; Department of Neurology, Changzheng Hospital, Naval Medical University, Shanghai, 200003, China. jianhuazh11@126.com.
Curr Med Sci ; 41(4): 705-711, 2021 Aug.
Article em En | MEDLINE | ID: mdl-34403095
ABSTRACT

OBJECTIVE:

To compare the performances among three different systems for video head impulse test (vHIT), and to identify an optimal target angle for precisely evaluating the function of vertical semicircular canals in vHIT.

METHODS:

A two-center prospective study was done. Participants were sit 1.2 m away from the wall in a noise-proved room that dedicated for vHIT experiments. During the comparison experiments, similar settings were ensured in both hospitals, with the same distance to wall and angle of staring. For each equipment, the procedures followed the developers' recommendations. The same examiner performed the comparison between two systems in one location. For the eye-position projects, targets were placed on the wall sequentially at the pre-marked lines for different angles. For the comparison projects, 9 and 13 participants were recruited, respectively. Any participant with otologic or vestibular disorders was excluded. A total of 26 healthy participants were recruited in the eye-position experiments, 16 of which were further involved in inter-examiner tests.

RESULTS:

Our evaluations of three different systems showed that a new vHIT system, VertiGoggles® ZT-VNG-I (VG) performed as good as the long-tested Otometrics® ICS impulse (Oto) and EyeSeeCam® (ESC). During the comparison, we validated 25-degree, instead of right ahead at 0 degree, is a better place to set the targets when torsion was applied at vertical semicircular canal planes.

CONCLUSION:

The new VG system is good for clinical practices. Furthermore, we proposed a new protocol to set the targets 25 degrees from right ahead after tilting head 45 degrees to evaluate vertical canals during vHIT.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Canais Semicirculares / Olho / Teste do Impulso da Cabeça Tipo de estudo: Guideline / Observational_studies Limite: Adult / Female / Humans / Male Idioma: En Revista: Curr Med Sci Ano de publicação: 2021 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Canais Semicirculares / Olho / Teste do Impulso da Cabeça Tipo de estudo: Guideline / Observational_studies Limite: Adult / Female / Humans / Male Idioma: En Revista: Curr Med Sci Ano de publicação: 2021 Tipo de documento: Article País de afiliação: China