Your browser doesn't support javascript.
loading
Proximal ascending aorta size is associated with the incidence of de novo aortic insufficiency with left ventricular assist device.
Nishida, Hidefumi; Song, Tae; Onsager, David; Nguyen, Ann; Grinstein, Jonathan; Chung, Bow; Smith, Bryan; Kalantari, Sara; Sarswat, Nitasha; Kim, Gene; Pinney, Sean; Jeevanandam, Valluvan; Ota, Takeyoshi.
Afiliação
  • Nishida H; Department of Surgery, Section of Cardiac Surgery, The University of Chicago Medicine, 5841S Maryland Avenue, MC5040, Chicago, IL, 60637, USA.
  • Song T; Department of Surgery, Section of Cardiac Surgery, The University of Chicago Medicine, 5841S Maryland Avenue, MC5040, Chicago, IL, 60637, USA.
  • Onsager D; Department of Surgery, Section of Cardiac Surgery, The University of Chicago Medicine, 5841S Maryland Avenue, MC5040, Chicago, IL, 60637, USA.
  • Nguyen A; Department of Medicine, The University of Chicago Medicine, Chicago, IL, USA.
  • Grinstein J; Department of Medicine, The University of Chicago Medicine, Chicago, IL, USA.
  • Chung B; Department of Medicine, The University of Chicago Medicine, Chicago, IL, USA.
  • Smith B; Department of Medicine, The University of Chicago Medicine, Chicago, IL, USA.
  • Kalantari S; Department of Medicine, The University of Chicago Medicine, Chicago, IL, USA.
  • Sarswat N; Department of Medicine, The University of Chicago Medicine, Chicago, IL, USA.
  • Kim G; Department of Medicine, The University of Chicago Medicine, Chicago, IL, USA.
  • Pinney S; Department of Medicine, The University of Chicago Medicine, Chicago, IL, USA.
  • Jeevanandam V; Department of Surgery, Section of Cardiac Surgery, The University of Chicago Medicine, 5841S Maryland Avenue, MC5040, Chicago, IL, 60637, USA.
  • Ota T; Department of Surgery, Section of Cardiac Surgery, The University of Chicago Medicine, 5841S Maryland Avenue, MC5040, Chicago, IL, 60637, USA. tota@surgery.bsd.uchicago.edu.
Heart Vessels ; 37(4): 647-653, 2022 Apr.
Article em En | MEDLINE | ID: mdl-34585275
ABSTRACT
We sought to assess the impact of the aortic root geometry on developing de novo aortic insufficiency (AI) in patients undergoing left ventricular assist device (LVAD). In total, 114 patients underwent LVAD implantation between February 2016 and January 2020 were included in this study (HeartMate3 N = 68, HeartWare N = 46). Significant aortic insufficiency was defined as mild-to-moderate or greater in echocardiography. The cohort was divided into two groups; those who developed significant AI (Group AI n = 13) and did not (Group non-AI n = 101). The primary outcomes of interest included late survival and predictors for significant AI. The patients in Group AI were older than Group non-AI (62.6 ± 11.9 vs 51.3 ± 14.0 years, p < 0.01). The diameter of proximal ascending aorta in Group AI was larger than Group non-AI (31.0 ± 5.0 vs 27.4 ± 4.3 mm, p < 0.01). Aortic valve remained closed in 53.8% in Group AI and 36.6% in Group non-AI (p = 0.24). The late survival was not significantly different between the groups (67.1% vs 76.0% at 3 years, log rank = 0.97). The Cox hazard model showed that larger proximal ascending aortic diameter/BSA (HR 1.55, CI 1.19-2.04, p < 0.01) and not-opening aortic valve (HR 4.73, CI 1.43-16.9, p = 0.01) were independent risk factors for significant AI. The cutoff value of proximal ascending aortic diameter/BSA was 15.5 (area under curve 0.770, sensitivity 0.69, specificity 0.79). Dilated proximal ascending aorta at the time of LVAD surgery and not-opening aortic valve during follow-up were associated with the incidence of de novo significant AI.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Insuficiência da Valva Aórtica / Coração Auxiliar Tipo de estudo: Diagnostic_studies / Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Heart Vessels Assunto da revista: CARDIOLOGIA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Insuficiência da Valva Aórtica / Coração Auxiliar Tipo de estudo: Diagnostic_studies / Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Heart Vessels Assunto da revista: CARDIOLOGIA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Estados Unidos