Effect of decreased contrast injection flow rate on aortic enhancement in 80-KV peak CT with contrast dose reduction.
Morisaka, Hiroyuki; Matsuura, Koichiro; Yamaguchi, Haruomi; Ichikawa, Tomoaki; Onishi, Hiroshi.
Afiliação
  • Morisaka H; Department of Radiology, 38146University of Yamanashi, Chuo, Yamanashi, Japan.
  • Matsuura K; Diagnostic Radiology, 183786Saitama Medical University International Medical Center, Hidaka, Saitama, Japan.
  • Yamaguchi H; Diagnostic Radiology, 183786Saitama Medical University International Medical Center, Hidaka, Saitama, Japan.
  • Ichikawa T; Department of Radiology, Graduate School of Medicine, 38551University of Tokyo, Bunkyo-ku, Tokyo, Japan.
  • Onishi H; Department of Diagnostic Radiology and Nuclear Medicine, 38357Gunma University Graduate School of Medicine, Maebashi, Gunma, Japan.
Acta Radiol ; 64(1): 353-359, 2023 Jan.
Article em En | MEDLINE | ID: mdl-34923851
ABSTRACT

BACKGROUND:

Effect of decreased injection flow rate of contrast agent at the same iodine dose and delivery rate on aortic enhancement has not been clearly elucidated.

PURPOSE:

To evaluate the effect of decreased injection flow rate of contrast agent on aortic peak enhancement in a dynamic flow phantom and on aortic enhancement in clinical dynamic 80-kVp computed tomography (CT) with contrast dose reduction. MATERIAL AND

METHODS:

In the dynamic flow phantom experiment, the effect of a decreased injection flow rate at the same total iodine dose and delivery rate on simulated aortic peak enhancement was evaluated. In the clinical retrospective study, we searched 312 patients with renal dysfunction who underwent an 80-kVp abdominal dynamic CT with 40% reduction of contrast agent from a standard 120-kVp protocol and measured the aortic enhancement at the level of the hepatic hilum. Independent predictors for aortic enhancement were determined by multiple linear regression analysis, and after adjustment of significant predictors, independent variables for acquiring optimal aortic enhancement, ≥300 HU, were determined by multiple logistic regression analysis.

RESULTS:

In the phantom experiment, decreased flow rate showed a significant but small descent effect (6%-9%) on simulated aortic peak enhancement. In the multiple linear regression analysis, only age was an independent predictor of aortic enhancement; there was no independent predictor for optimal age-adjusted aortic enhancement of ≥300 HU.

CONCLUSIONS:

Decreased injection flow rate had a small influence on aortic enhancement in vitro but had no significant effect on the aortic enhancement in clinical dynamic 80-kVp CT.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Meios de Contraste / Iodo Tipo de estudo: Guideline / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Acta Radiol Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Meios de Contraste / Iodo Tipo de estudo: Guideline / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Acta Radiol Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Japão