Your browser doesn't support javascript.
loading
Characterization and effectiveness of a violet LED light for in-office whitening.
Kury, Matheus; Rueggeberg, Frederick A; Soto-Montero, Jorge R; André, Carolina B; Resende, Bruna A; Giannini, Marcelo; Cavalli, Vanessa.
Afiliação
  • Kury M; Department of Restorative Dentistry, Piracicaba Dental School, University of Campinas, 901 Limeira Avenue, Areião, Piracicaba, SP, 13414-903, Brazil.
  • Rueggeberg FA; Department of Restorative Sciences, Dental College of Georgia, Augusta University, Augusta, GA, USA.
  • Soto-Montero JR; Department of Restorative Dentistry, Piracicaba Dental School, University of Campinas, 901 Limeira Avenue, Areião, Piracicaba, SP, 13414-903, Brazil.
  • André CB; Department of Restorative Dentistry, Dental School, Universidad de Costa Rica, San José, SP, Costa Rica.
  • Resende BA; Department of Restorative Dentistry, Piracicaba Dental School, University of Campinas, 901 Limeira Avenue, Areião, Piracicaba, SP, 13414-903, Brazil.
  • Giannini M; Department of Restorative Dentistry, Dental School, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil.
  • Cavalli V; Department of Restorative Dentistry, Piracicaba Dental School, University of Campinas, 901 Limeira Avenue, Areião, Piracicaba, SP, 13414-903, Brazil.
Clin Oral Investig ; 26(5): 3899-3910, 2022 May.
Article em En | MEDLINE | ID: mdl-35006294
ABSTRACT

OBJECTIVES:

This study characterized a violet LED light (V-LED; bright max whitening) tooth whitening device and evaluated its efficacy on stained enamel compared to hydrogen peroxide (HP). MATERIALS AND

METHODS:

Characterization of the V-LED beam profile was performed using a laser beam-profiler. The irradiance was measured throughout an exposure cycle at 0- and 8-mm distances using an integrating sphere and a spectral radiometer. Bovine enamel/dentin blocks stained with black tea (BT), cigarette smoke (CS), or without staining (CONT) were subjected to V-LED or 40% HP (n = 10/group). Color parameters (ΔL, Δa, Δb, and ΔE00) were measured using a digital spectrophotometer. Light transmission was estimated through 1-mm-thick bovine enamel slices (n = 5). ΔL, Δb, ΔE00, and irradiance were analyzed by two-way ANOVAs and Tukey's tests, Δa by Kruskal-Wallis and Mann-Whitney tests, and light transmission by t-test (α = 5%).

RESULTS:

Heterogeneous beam distribution was observed for the emitting V-LED chips. After 20 sequential exposures, irradiance levels were reduced 25-50%, regardless of the distance from V-LED. Localized irradiance values were statistically different between beam locations and different distances from the target. V-LED produced lower ΔE00, ΔL, Δa, and Δb values than HP for CONT and BT, with no differences for CS. Light transmittance decreased approximately 98% through 1-mm thick enamel.

CONCLUSIONS:

V-LED irradiance was heterogeneous and decreased throughout the exposure cycles and was also greatly reduced with increasing tip distance. V-LED produced a significantly lower whitening effect on BT and control teeth. CLINICAL RELEVANCE This study contributes to the knowledge of V-LED and its clinical use.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Clareamento Dental / Clareadores Dentários Limite: Animals Idioma: En Revista: Clin Oral Investig Assunto da revista: ODONTOLOGIA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Brasil

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Clareamento Dental / Clareadores Dentários Limite: Animals Idioma: En Revista: Clin Oral Investig Assunto da revista: ODONTOLOGIA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Brasil