Focused ultrasound neuromodulation on a multiwell MEA.
Bioelectron Med
; 8(1): 2, 2022 Jan 27.
Article
em En
| MEDLINE
| ID: mdl-35081966
BACKGROUND: Microelectrode arrays (MEA) enable the measurement and stimulation of the electrical activity of cultured cells. The integration of other neuromodulation methods will significantly enhance the application range of MEAs to study their effects on neurons. A neuromodulation method that is recently gaining more attention is focused ultrasound neuromodulation (FUS), which has the potential to treat neurological disorders reversibly and precisely. METHODS: In this work, we present the integration of a focused ultrasound delivery system with a multiwell MEA plate. RESULTS: The ultrasound delivery system was characterised by ultrasound pressure measurements, and the integration with the MEA plate was modelled with finite-element simulations of acoustic field parameters. The results of the simulations were validated with experimental visualisation of the ultrasound field with Schlieren imaging. In addition, the system was tested on a murine primary hippocampal neuron culture, showing that ultrasound can influence the activity of the neurons. CONCLUSIONS: Our system was demonstrated to be suitable for studying the effect of focused ultrasound on neuronal cultures. The system allows reproducible experiments across the wells due to its robustness and simplicity of operation.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
En
Revista:
Bioelectron Med
Ano de publicação:
2022
Tipo de documento:
Article
País de afiliação:
Holanda