Your browser doesn't support javascript.
loading
Release of metal elements from the Essure implant: A prospective cohort study.
Parant, François; Bost, Muriel; Nohuz, Erdogan; Prost, Julie; Cerruto, Emanuele; Moret, Stephanie; Maurin, Julien; Chene, Gautier.
Afiliação
  • Parant F; Trace Element Analysis Laboratory, Biochemistry Department, CBAPS, Centre hospitalier Lyon-Sud, Chemin du Grand-Revoyet, University hospital of Lyon, 69495 Pierre-Bénite, France.
  • Bost M; Trace Element Analysis Laboratory, Biochemistry Department, CBAPS, Centre hospitalier Lyon-Sud, Chemin du Grand-Revoyet, University hospital of Lyon, 69495 Pierre-Bénite, France.
  • Nohuz E; Department of Gynecology, Hôpital Femme Mère Enfant, HFME, 59 boulevard Pinel, University Hospital of Lyon, 69500 Bron, France.
  • Prost J; Trace Element Analysis Laboratory, Biochemistry Department, CBAPS, Centre hospitalier Lyon-Sud, Chemin du Grand-Revoyet, University hospital of Lyon, 69495 Pierre-Bénite, France.
  • Cerruto E; Department of Gynecology, Hôpital Femme Mère Enfant, HFME, 59 boulevard Pinel, University Hospital of Lyon, 69500 Bron, France.
  • Moret S; Department of Gynecology, Hôpital Femme Mère Enfant, HFME, 59 boulevard Pinel, University Hospital of Lyon, 69500 Bron, France.
  • Maurin J; Trace Element Analysis Laboratory, Biochemistry Department, CBAPS, Centre hospitalier Lyon-Sud, Chemin du Grand-Revoyet, University hospital of Lyon, 69495 Pierre-Bénite, France.
  • Chene G; Department of Gynecology, Hôpital Femme Mère Enfant, HFME, 59 boulevard Pinel, University Hospital of Lyon, 69500 Bron, France; University Claude Bernard of Lyon 1, EMR 3738 CICLY, 69000 Lyon, France. Electronic address: chenegautier@yahoo.fr.
Eur J Obstet Gynecol Reprod Biol ; 273: 20-25, 2022 Jun.
Article em En | MEDLINE | ID: mdl-35447464
ABSTRACT
OBJECTIVE(S) The causal mechanistic relationships between Essure® and adverse effects are unclear, but corrosion in the in-vivo environment with release of metal ions may be suspected. Here we evaluated the concentrations of nickel (Ni), chromium (Cr) and tin (Sn) in the peritoneal fluid (PF) and in the fallopian tube (FT) during laparoscopic Essure® removal compared to a control group. STUDY

DESIGN:

Ni, Cr and Sn concentrations were determined in the PF and FT from two groups(group A symptomatic patients with Essure®) vs group B (control group without Essure®) by Inductively Coupled Plasma Mass Spectrometry analysis. Correlation between metal elements concentrations and reported pre-operative symptoms was also investigated.

RESULTS:

There were 131 patients in group A vs 92 control patients in group B. The concentrations of Cr and Ni in PF between both groups were significantly different (p < 0.0001) while there was no statistical difference for Sn (p = 0.58). There was also a significantly higher concentration in the FT for the 3 metal elements in group A than in group B (p < 0.0001). There were differential dynamics of the levels of metal elements based on the length of time between the placement and removal of Essure®.

CONCLUSIONS:

There was a chronic exposure to metal elements in symptomatic patients with Essure® raising the question of the relationship between adverse effects and these potential toxic metals.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Esterilização Tubária / Laparoscopia Tipo de estudo: Etiology_studies / Observational_studies Limite: Female / Humans Idioma: En Revista: Eur J Obstet Gynecol Reprod Biol Ano de publicação: 2022 Tipo de documento: Article País de afiliação: França

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Esterilização Tubária / Laparoscopia Tipo de estudo: Etiology_studies / Observational_studies Limite: Female / Humans Idioma: En Revista: Eur J Obstet Gynecol Reprod Biol Ano de publicação: 2022 Tipo de documento: Article País de afiliação: França