Your browser doesn't support javascript.
loading
Light-initiated chemiluminescent assay of 17ß-estradiol metrological traceability system established by manufacturer according to ISO17511:2020 and basic performance evaluation performed by clinical end-users.
Shang, Chenyu; Yuan, Xue; Lin, Haibiao; Liu, Dongdong; Yan, Xiaoxin; Ren, Xinxin; Lin, Xiyang; Di, Huang; Li, Huiqiang.
Afiliação
  • Shang C; Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China.
  • Yuan X; The Second Hospital of Tianjin Medical University, Tianjin, China.
  • Lin H; Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China.
  • Liu D; Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China.
  • Yan X; Standardization & Performance Evaluation Laboratory, Chemclin Diagnostics Co. Ltd, Beijing, China.
  • Ren X; Standardization & Performance Evaluation Laboratory, Chemclin Diagnostics Co. Ltd, Beijing, China.
  • Lin X; Standardization & Performance Evaluation Laboratory, Chemclin Diagnostics Co. Ltd, Beijing, China.
  • Di H; Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China.
  • Li H; Department of Clinical Immunology, School of Medical Laboratory, Tianjin Medical University, Tianjin, China.
J Clin Lab Anal ; 36(6): e24436, 2022 Jun.
Article em En | MEDLINE | ID: mdl-35471729
ABSTRACT

BACKGROUND:

In order to ensure the accuracy of the product, we established 1st model of metrological traceability hierarchy for light-initiated chemiluminescent assay (LICA) of 17ß-estradiol (E2 ) at the manufacturer, based on International Organization for Standardization (ISO) 175112020. Moreover, we verified/validated the basic performance (such as matrix effect and long-term stability of end-user IVD MD calibrator, precision, linearity interval, accuracy/ trueness, and detection capability) at the clinical end-user.

METHODS:

Human serum samples were used in this study. E2 was detected by mass spectrometry (MS) and LICA. The metrological traceability of LICA for E2 was established according to ISO 17511 2020 standards, and pools of human samples were used as the m.3. secondary calibrator. Precision was validated according to Clinical and Laboratory Standards Institute (CLSI) EP05-A3. The linear interval was verified according to CLSI EP06-ED2. Comparison of accuracy and trueness of E2 with MS and Roche according to CLSI EP09-A3. The detection capability was validated according to EP17-A2. Matrix effect and long-term stability evaluation of end-user IVD MD calibrator were carried out according to CLSI EP14-A2, EP25-A. Statistical software was used for data analyses.

RESULTS:

The use of pools of human samples and fine adjusting calibrators ensured the accuracy of end-user test results. The metrological traceability of LICA for E2 was established. It showed excellent precision, meeting the requirements of allowable imprecision (7.5%). The allowable deviation from linearity (ADL) of 5% was allowed to show a good linear interval (12.52-4167.25 pg/ml). The accuracy/ trueness was verified, and relative deviation in the medical decision level met the performance specification of 10.03% compared with MS or Roche. The validated limit of blank, limit of detection, and limit of quantitation of E2 were 4.95 pg/ml, 8.93 pg/ml, and 9.88 pg/ml, respectively (the allowed imprecision is 20.00%). The interference rate of E2 ranged from -5.5% to 6.6%.

CONCLUSION:

LICA showed high sensitivity, high specificity, excellent precision, wide linearity interval, IVD MD calibrator has long-term stability, and no matrix effect. The metrological traceability of E2 established by using pools of human samples as M.3. can deliver accuracy to the end-user IVD MD and show good consistency with MS and Roche.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Serviços de Laboratório Clínico / Medições Luminescentes Tipo de estudo: Guideline / Prognostic_studies Limite: Humans Idioma: En Revista: J Clin Lab Anal Assunto da revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2022 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Serviços de Laboratório Clínico / Medições Luminescentes Tipo de estudo: Guideline / Prognostic_studies Limite: Humans Idioma: En Revista: J Clin Lab Anal Assunto da revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2022 Tipo de documento: Article País de afiliação: China