Finite element analysis of the stump-ischial containment socket interaction: a technical note.
Fougeron, Nolwenn; Rohan, Pierre-Yves; Rose, Jean-Loïc; Bonnet, Xavier; Pillet, Hélène.
Afiliação
  • Fougeron N; Institut de Biomécanique Humaine Georges Charpak, Arts et Métiers ParisTech, 151 bd de l'Hôpital, 75013, Paris; Proteor, Recherche et développement, 5 boulevard Winston Churchill, 21000 Dijon, France. Electronic address: nolwenn.fougeron@univ-grenoble-alpes.fr.
  • Rohan PY; Institut de Biomécanique Humaine Georges Charpak, Arts et Métiers ParisTech, 151 bd de l'Hôpital, 75013, Paris.
  • Rose JL; Proteor, Recherche et développement, 5 boulevard Winston Churchill, 21000 Dijon, France.
  • Bonnet X; Institut de Biomécanique Humaine Georges Charpak, Arts et Métiers ParisTech, 151 bd de l'Hôpital, 75013, Paris.
  • Pillet H; Institut de Biomécanique Humaine Georges Charpak, Arts et Métiers ParisTech, 151 bd de l'Hôpital, 75013, Paris. Electronic address: helene.pillet@ensam.eu.
Med Eng Phys ; 105: 103829, 2022 07.
Article em En | MEDLINE | ID: mdl-35781389
ABSTRACT
The role of the above-knee socket is to ensure the load transfer via the coupling of residual limb-prosthesis with minimal discomfort and without damaging the soft tissues. Modelling is a potential tool to predict socket fit prior to manufacture. However, state-of-the-art models only include the femur in soft tissues submitted to static loads neglecting the contribution of the hip joint. The hip joint is particularly challenging to model because it requires to compute the forces of muscles inserting on the residual limb. This work proposes a modelling of the hip joint including the estimation of muscular forces using a combined MusculoSKeletal (MSK)/Finite Element (FE) framework. An experimental-numerical approach was conducted on one femoral amputee subject. This allowed to i) model the hip joint and personalise muscular forces, ii) study the impact of the ischial support, and iii) evaluate the interface pressure. A reduction of the gluteus medius force from the MSK modelling was noticed when considering the ischial support. Interface pressure, predicted between 63 to 71 kPa, agreed with experimental literature data. The contribution of the hip joint is a key element of the modelling approach for the prediction of the socket interface pressure with the residual limb soft tissues.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Articulação do Quadril / Cotos de Amputação Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Med Eng Phys Assunto da revista: BIOFISICA / ENGENHARIA BIOMEDICA Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Articulação do Quadril / Cotos de Amputação Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Med Eng Phys Assunto da revista: BIOFISICA / ENGENHARIA BIOMEDICA Ano de publicação: 2022 Tipo de documento: Article