Wireless Communication Between Paired Leadless Pacemakers for Dual-Chamber Synchrony.
Cantillon, Daniel J; Gambhir, Alok; Banker, Rajesh; Rashtian, Mayer; Doshi, Rahul; Badie, Nima; Booth, Daniel; Yang, Weiqun; Nee, Peter; Fishler, Matthew; Ligon, David; Neuzil, Petr; Knops, Reinoud E.
Afiliação
  • Cantillon DJ; Cleveland Clinic, Heart and Vascular Institute, OH (D.J.C.).
  • Gambhir A; Northside Hospital Cardiovascular Institute, Sandy Springs, GA (A.G.).
  • Banker R; Hoag Hospital, Newport Beach (R.B.).
  • Rashtian M; Huntington Hospital, Pasadena, CA (M.R.).
  • Doshi R; HonorHealth Research Institute, Scottsdale, AZ (R.D.).
  • Badie N; Abbott, Sunnyvale, CA (N.B., D.B., W.Y., P.N., M.F., D.L.).
  • Booth D; Abbott, Sunnyvale, CA (N.B., D.B., W.Y., P.N., M.F., D.L.).
  • Yang W; Abbott, Sunnyvale, CA (N.B., D.B., W.Y., P.N., M.F., D.L.).
  • Nee P; Abbott, Sunnyvale, CA (N.B., D.B., W.Y., P.N., M.F., D.L.).
  • Fishler M; Abbott, Sunnyvale, CA (N.B., D.B., W.Y., P.N., M.F., D.L.).
  • Ligon D; Abbott, Sunnyvale, CA (N.B., D.B., W.Y., P.N., M.F., D.L.).
  • Neuzil P; Na Homolce Hospital, Prague, Czech Republic (P.N.).
  • Knops RE; Academic Medical Center, Amsterdam, the Netherlands (R.E.K.).
Circ Arrhythm Electrophysiol ; 15(7): e010909, 2022 07.
Article em En | MEDLINE | ID: mdl-35861976
ABSTRACT

BACKGROUND:

Leadless pacemakers (LPs) can mitigate conventional pacemaker complications related to the transvenous leads and subcutaneous pocket surrounding the pulse generator. Although single-chamber leadless pacing has been established, multichamber pacing requires wireless bidirectional communication across multiple LPs to maintain synchrony. This preclinical study demonstrates the chronic performance of implant-to-implant (i2i) communication that achieves synchronous, dual-chamber pacing with 2 LPs.

METHODS:

The i2i communication modality employs subthreshold electrical signals conducted between implanted LPs through the blood and myocardial tissue on a beat-by-beat basis. Right atrial and right ventricular LPs were implanted in 9 ovine subjects. The i2i transmission performance was evaluated 13 weeks after implant.

RESULTS:

Right atrial and right ventricular LPs were implanted successfully and without complication in 9 ovine subjects. A total of 8715±457 right atrial-to-right ventricular and right ventricular-to-right atrial transmissions were sent per hour, with a success rate of 99.2±0.9%. Of periods with i2i communication failure when DDD pacing was not possible, 97.3±1.8% were resolved within 6 s.

CONCLUSIONS:

For the first time, synchronized, dual-chamber pacing has been demonstrated in a chronic preclinical feasibility study by 2 leadless pacemakers using beat-to-beat, wireless communication, achieving a success rate of 99.2%.