Evaluation of a Wearable in-Ear Sensor for Temperature and Heart Rate Monitoring: A Pilot Study.
Ellebrecht, David Benjamin; Gola, Damian; Kaschwich, Mark.
Afiliação
  • Ellebrecht DB; Department of Surgery, University Medical Center Schleswig-Holstein, Campus Luebeck, Ratzeburger Allee 160, 23538, Luebeck, Germany. D.Ellebrecht@LungenClinic.de.
  • Gola D; Department of Thoracic Surgery, LungenClinic Großhansdorf, Woehrendamm 80, 22927, Grosshansdorf, Germany. D.Ellebrecht@LungenClinic.de.
  • Kaschwich M; Institute of Medical Biometry and Statistics, University of Lübeck, Ratzeburger Allee 160, 23562, Luebeck, Germany.
J Med Syst ; 46(12): 91, 2022 Nov 04.
Article em En | MEDLINE | ID: mdl-36329338
In the context of the COVID-19 pandemic, wearable sensors are important for early detection of critical illness especially in COVID-19 outpatients. We sought to determine in this pilot study whether a wearable in-ear sensor for continuous body temperature and heart rate monitoring (Cosinuss company, Munich) is sufficiently accurate for body temperature and heart rate monitoring. Comparing with several anesthesiologic standard of care monitoring devices (urinary bladder and zero-heat flux thermometer and ECG), we evaluated the in-ear sensor during non-cardiac surgery (German Clinical Trials Register Reg.-No: DRKS00012848). Limits of Agreement (LoA) based on Bland-Altman analysis were used to study the agreement between the in-ear sensor and the reference methods. The estimated LoA of the Cosinuss One and bladder temperature monitoring were [-0.79, 0.49] °C (95% confidence intervals [-1.03, -0.65] (lower LoA) and [0.35, 0.73] (upper LoA)), and [-0.78, 0.34] °C (95% confidence intervals [-1.18, -0.59] (lower LoA) and [0.16, 0.74] (upper LoA)) of the Cosinuss One and zero-heat flux temperature monitoring. 89% and 79% of Cosinuss One temperature monitoring were within ± 0.5 °C limit of bladder and zero-heat flux monitoring, respectively. The estimated LoA of Cosinuss One and ECG heart rate monitoring were [-4.81, 4.27] BPM (95% confidence intervals [-5.09, -4.56] (lower LoA) and [4.01, 4.54] (upper LoA)). The proportion of detection differences within ± 2BPM was 84%. Body temperature and heart rate were reliably measured by the wearable in-ear sensor.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Dispositivos Eletrônicos Vestíveis / COVID-19 Tipo de estudo: Diagnostic_studies / Screening_studies Limite: Humans Idioma: En Revista: J Med Syst Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Dispositivos Eletrônicos Vestíveis / COVID-19 Tipo de estudo: Diagnostic_studies / Screening_studies Limite: Humans Idioma: En Revista: J Med Syst Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Alemanha